Press button      to switch on the device. Press the sensor against the 
surface. The sensor must be kept strictly perpendicular to the surface. 

The result of coating thickness measurement will appear on the display. 

If the device is in the single measurement mode, carry out the next 

measurement three seconds after removing the device from the surface.

STATISTICS

Short press button  

ZERO/CAL

  to open the menu.  

Press buttons         and        to select 

“COUNT”(Statistics)

 and press 

buttons  

OK

The device saves statistics on the last 80 measurements:

Num – the number of taken measurements (up to 80)

Max – maximal measurement value

Min – minimal measurement value

Mean – average value

Sdev – square deviation

If the number of measurements exceeds 80, new measurements will be 

saved over the previous ones.

MEMORY

Short press button  

ZERO/CAL

 to open the menu.

Press buttons         and        to select 

“RECALL” (Memory)

 and press button 

OK

Press buttons         and       , to view the results of taken measurements.
To exit the menu press button  

ESC

  required number of times.

MEMORY CLEANING

Short press button  

ZERO/CAL 

to open the menu.

 

Pres buttons             and        to select 

“DELETE” 

and 

press button 

OK

. Select the measurement to be deleted and press button 

OK

. Measurement result will 

be deleted. To exit the menu press button 

ESC

 required number of times.

RESET TO FACTORY SETTINGS AND CALIBRATION

This function is used for the purpose of calibration. Prior to making measurements on a magnetic 
surface carry out calibration on iron plate included in the delivery package, or on aluminum plate 
before making measurements on a non-magnetic surface.
Short press button 

ZERO/CAL 

to open the menu. Press buttons          and         to select 

“RECOVERY” 

(reset to factory settings)

 and press button 

OK

Press buttons       and      to select 

«Yes»

 and press button 

OK

 to confirm. The device will switch off 

automatically.
Press button         to switch it on. Zero calibration is finished. 
5 seconds after switching on the device press button 

ZERO/CAL

. The device will switch off 

automatically. Press button           to switch it on again.

Prepare film reference samples 50µm, 100µm, 250µm, 500µm и 1000µm.

Press button 

ZERO/CAL 

to open the menu and select 

“CALIBRATION”

.

Press buttons        and        , select 

«0»

 and press button 

OK

 to confirm. 

Take the metal plate, put the sensor down onto the plare and do not detach the sensor from its surface 
for 3 seconds. Press button         .
Press button  

OK

“Success”

 will appear on the display. Press button 

ESC

 to return to previous section.

Press buttons         and       , select 

«50»

 and press button 

OK

 to confirm. 

Place the film 50µm on a metal plate. Bring the sensor down onto the surface and do not detach the 
sensor from the surface for 3 seconds. Press button          .
Press button  

OK

“Success”

 will appear on the display. Press button 

ESC

  to return to previous section.

Thus, go through other calibration points using films 100µm, 250µm, 500µm и 1000µm included in 

the delivery package.

Attention:

 It is recommended to make calibration at the mode of automatic probe and continuous 

measurement.

DATA TRANSFER VIA USB INTERFACE

Connect the device with the PC or laptop by using of the USB cable included in the delivery package. 
The program for transferring data to the PC is located on the CD included in the delivery package. 
Run the program. Press the button 

OK

. If the connection between the device and the program is 

successful, a USB symbol will appear in the upper right corner of the display.
Press button                                                           in the program. 
The measuring results will be displayed in the program. 

Attention:

 the program displays only the results of group measurements.

Data downloading

CARE AND MAINTENANCE

Attention! This device is a precision instrument and requires careful handling. Maintenance of the 

following recommendations will extend the life of the device:

- Protect the device from bumps, falling and intense vibration; do not allow moisture, dust and foreign 

objects get inside the device. 

- If liquids get inside the device, first remove the batteries, then contact a service center.

- Do not store and use the device in increased humidity conditions. 

-  Do  not  store  the  product  in  rooms  where  temperature  is  below  -10°С.    After  storage  in  low 

temperature conditions and subsequent transfer to a warm room, the device heats up, causing 

moisture condense inside the device and damage microcircuits.

- Do not expose the device to direct sunlight and protect it from prolonged exposure to sunlight and 

high temperature.

- Clean the device with a soft cloth slightly made damp with a mild soap solution. Do not use cleaning 

solvents or abrasives. 

Failure to observe the following rules may lead to electrolyte leakage from the batteries and 

damage the device:

-  Remove the batteries from the device if you don’t use it for a long time.

- Do not use batteries of different types, with different charge level.

- Do not leave discharged batteries in the device.

UTILIZATION

Expired tools, accessories and package should be passed for waste recycle. Please send the product to 

the following address for proper recycle:

CONDTROL GmbH 

Wasserburger Strasse 9 

84427 Sankt Wolfgang 

Germany 

Do not throw the product in municipal waste!

According to European directive 2002/96/ЕC expired measuring tools and their components must be 

collected separately and submitted to environmentally friendly recycle of wastes.

WARRANTY

All CONDTROL GmbH products go through post-production control and are governed by the following 

warranty terms. The buyer’s right to claim about defects and general provisions of the current 

legislation do not expire.
1) CONDTROL GmbH agrees to eliminate all defects in the product, discovered while warranty period, 

that represent the defect in material or workmanship in full volume and at its own expense.
2) The warranty period is 24 months and starts from the date of purchase by the end customer (see 

the original supporting document).
3) The warranty doesn’t cover defects resulting from wear and tear or improper use, malfunction of 

the product caused by failure to observe the instructions of this user manual, untimely maintenance 

and service and insufficient care, the use of non-original accessories and spare parts. Modifications in 

design of the product relieve the seller from responsibility for warranty works. The warranty does not 

cover cosmetic damage, that doesn’t hinder normal operation of the product.
4) CONDTROL GmbH reserves the right to decide on replacement or repair of the device. 
5) Other claims not mentioned above, are not covered by the warranty. 
6) After holding warranty works by CONDTROL GmbH warranty period is not renewed or extended.
7) CONDTROL GmbH is not liable for loss of profit or inconvenience associated with a defect of the 

device, rental cost of alternative equipment for the period of repair.   
This warranty applies to German law except provision of the United Nations Convention on contracts 

for the international sale of goods (CISG).

In warranty case please return the product to retail seller or send it with description of defect to the 

following address:

CONDTROL GmbH 

Wasserburger Strasse 9 

84427 Sankt Wolfgang 

Germany

Summary of Contents for Paint PRO

Page 1: ...on the battery cover and then remove it Install the batteries observing polarity Put the battery cover back and use a screwdriver to tighten the screw Use alkaline batteries only both batteries must b...

Page 2: ...lts will be displayed in the program Attention the program displays only the results of group measurements Data downloading CARE AND MAINTENANCE Attention This device is a precision instrument and req...

Page 3: ...ppe abzudrehen dann nehmen Sie sie ab Achten Sie dabei auf die richtige Polung Stellen Sie die Abdeckklappe zur ck und ziehen Sie die Schrauben mit Hilfe des Schraubendrehers an Setzen Sie nur Alkali...

Page 4: ...gen angezeigt Data downloading PFLEGE UND BEDIENUNG Vorsicht Dieses Instrument ist ein pr zises Ger t und muss stets vorsichtig behandelt werden Die Beachtung der folgenden Vorschriften verl ngert die...

Page 5: ...5 Language 7 Recovery 0 1500 1 0 999 0 01 1000 3 2 0 5 1 5 25 6 V 3 2 x 1 5 AAA 0 450 C 0 90 RH 10 500 C 0 90 RH 112 51 28 80 1 2 5 1 5 AAA 2 USB 1 1 CD 1 1 Self Test 1 4 6 10 8 9 3 5 7 2 2 3 4 5 6 7...

Page 6: ...OK ESC ZERO CAL DELETE OK OK ESC ZERO CAL RECOVERY OK Yes OK 5 ZERO CAL 50 m 100 m 250 m 500 m 1000 m ZERO CAL CALIBRATION 0 OK 3 OK Success ESC 50 OK 50 m 3 OK Success ESC 100 m 250 m 500 m 1000 m D...

Page 7: ...omoc wkr taka poluzowa ruby pokrywy baterii i zdj pokryw baterii Baterie nale y instalowa zgodnie z biegunowo ci Za o y pokryw baterii i dokr ci ruby rubokr tem U ywaj tylko baterii alkalicznych obie...

Page 8: ...K na urz dzeniu Je li po czenie pomi dzy urz dzeniem a programem zosta o pomy lnie nawi zane w prawym g rnym rogu wy wietlacza pojawi si nast puj cy symbol USB W programie nacisn na przycisk Za adowa...

Reviews: