background image

 

OMM Series 

 

User Manual 

 

 

Document No.: UM-OMM-2  Rev. A 
File Name: OMM Series Optical Multimeter User Guide 
Draft

 

Confidential

 

 

 

i

 

Disclaimer 

Copyright © 2016–2017 © ComSonics Inc. All rights reserved. 

No part of this document may be reproduced, stored in a retrieval system, or transmitted, in any form or 

by  any  means,  electronic,  mechanical,  photographic,  magnetic,  or  otherwise,  without  the  prior  written 

permission of ComSonics, Inc. 

ComSonics reserves the right to alter the equipment specifications and descriptions in this publication 

without prior notice. No part of this publication shall be deemed to be part of any contract or warranty 

unless specifically incorporated by reference into such contract or warranty. The information contained 

herein is merely descriptive in nature and does not constitute a binding offer for sale of the product 

described herein. ComSonics assumes no responsibility or liability arising from the use of the products 

described herein, except as expressly agreed to in writing by ComSonics. The use and purchase of this 

product do not convey a license under any patent rights, copyrights, trademark rights, or any intellectual 

property rights of ComSonics. Nothing hereunder constitutes a representation or warranty that using any 

products in the manner described herein will not infringe any patents of third parties. 

(The exclusion of implied warranties may not apply in all cases under some statutes, and thus the above 

exclusion may not  apply.) 

Specifications  are  subject  to  change  without  notice.  Every  attempt  has  been  made  to  make  this  manual 

complete, accurate, and up-to-date. Readers are cautioned, however, that ComSonics reserves the right to 

make changes without notice and shall not be responsible for any damages, including indirect, incidental or 

consequential  damages,  caused  by  reliance  on  the  material  presented,  including,  but  not  limited  to, 

omissions, typographical, arithmetical, or listing errors in the material. 

 

 

 

 

When you work with this equipment, take the following precautions: 

Use of controls or adjustments or performance of procedures other than those specified herein may result 

in hazardous radiation exposure. 

Never look at an optical fiber using a direct magnification fiber scope without verifying that all laser 

sources on the fiber are turned off. 

Never look directly into an optical connector. 

Install protective covers or caps over connectors when they are not being used. Always assume that 

exposed optical connectors and cables are energized. 

Always make an optical power measurement of the fiber or connector before cleaning to ensure the cable 

and connector are not energized. 

 

 

Summary of Contents for FHM2A01

Page 1: ...User Manual Optical Multimeter O OM MM M 2 2A A1 1 O OM MM M 2 2B B1 1 O OM MM M 2 2A A2 2 a an nd d O OM MM M 2 2B B2 2 R RE EV V A A J Ju ul ly y 2 20 01 18 8...

Page 2: ...resentation or warranty that using any products in the manner described herein will not infringe any patents of third parties The exclusion of implied warranties may not apply in all cases under some...

Page 3: ...symbol calls your attention to information that if ignored can adversely affect the performance of your ComSonics product or that can make a procedure needlessly difficult LASER DANGER The Laser symbo...

Page 4: ...AND FUNCTION OF COMPONENTS 1 1 1 INVENTORY 1 1 2 FEATURES 1 1 3 DESCRIPTION 1 CHAPTER 2 OPERATION 5 2 1 PERFORM OPTICAL POWER TEST 5 2 2 TWIN MODE 5 2 3 PERFORMING A FIBER LOSS TEST 6 CHAPTER 3 PC CON...

Page 5: ...1 2 Features A High performance Optical Power Meter B Interchangeable connectors C Measures six different wavelengths D Laser light source output on two or three wavelengths E Auto off function saves...

Page 6: ...nt reference setting 3 Modulation Indicator Displays modulated CW signal If none of the tones are shown on the display then they are all off 4 TWIN mode indicator a Single Auto wavelength detection of...

Page 7: ...ta files at once press and hold for 2 seconds The meter must be in the load mode where the stored measurements are shown for this function to work 4 Load key a Short press to enter Load mode Displays...

Page 8: ...hort press until beep sounds to power on with auto off function Active b Long press 2 Seconds after beep when powering on Disables auto off function c Meter will display auto off when function is enab...

Page 9: ...at bottom center of display 8 Turn off the meter 9 Disconnect the tested fiber clean connections and reconnect to fiber network 2 2 TWIN Mode When the ComSonics OLS in the multimeter is paired with a...

Page 10: ...B This provides a Reference level that will allow the power meter to measure signal level differences when measuring fiber loss 2 Repeat these steps for all wavelengths to be tested while ensuring the...

Page 11: ...r NOTE The first time an OMM series meter is connected a USB driver will be installed 3 A USB logo will be displayed on the meter LCD 4 If desired the company name user name and notes may be entered i...

Page 12: ...geable batteries and an AC charging adapter are supplied with the meter The rechargeable batteries may be replaced with standard 1 5 VDC AA batteries To replace the batteries 1 Use a common screwdrive...

Page 13: ...0dBm Measuring Range 70 to 10dBm 50 to 26dBm 70 to 10dBm 50 to 26dBm Laser Source Parameters Output Wavelength 1310 1550nm 1310 1490 1550nm Connector fixed FC PC interchangeable FC SC ST customized Mo...