8
T
Table 2: HTT-1S front panel Controls, Indicators, and Connectors
Item
Name
Function
1
Return Banana
Receptacle
Connected to the Return Lead on the Hipot or Ground Continuity Tester,
which is normally connected to dead metal parts of the equipment undergoing
dielectric test.
2
Output Cable
Connected to the high voltage output of the Hipot Tester being checked.
3
Arc View
When Dielectric Breakdown function is being checked, a spark will be
visually demonstrated the breakdown occurred.
4
Ground Test
Switch
When set to FAIL, an open ground is simulated and the tester should stop
testing and indicate a failure. When set to PASS, a normal passing result is
simulated.
5
Leakage Test
Switch
When set to FAIL, an excess leakage condition is simulated and the tester
should stop testing and indicate a failure. When set to PASS, a normal
passing result is simulated. Used for Dielectric Testers only.
6
Hipot Test
Switch
When set to FAIL, a dielectric breakdown will be simulated and a spark can
be seen in the Arc View window. When set to PASS, a normal passing result
is simulated. Used for Dielectric Testers only.
7
OFFSET
switch
Switch used to offset the HTT-1S cables for proper ground reading. Before
any testing is done, be sure to Offset the HTT-1S ground on the Hipot tester,
refer to setup procedure for offset instructions.