Section 1 

Operating Instructions 

 

Theory of Operation 

 
  The  HTT-1S  is  designed  to  be  a  solution  for  fast  and easy function testing of 
the  HT-S  series  of  dielectric  withstand  and  continuity  testers,  including  high 
current  models  delivering  up  to  30  amps.    It  allows  function  tests  of  all  three 
failure modes to ensure that your tester is functioning properly.   
  The HTT-1S is designed to verify calibration points at a fixed value. 
 
  The HTT-1S, used daily, will prove function of the safety testers before shifts 
or  before  testing  of  a  lot  of  products.    It  is  provided  with  three  switches  which 
simulate  passing  and  failing  equipment  for  ground  continuity,  high  leakage 
between  primary  circuits  and  ground,  and  dielectric  withstand  failure  between 
primary  and ground circuits.  While all three switches can be simultaneously set 
for  failing  modes,  we  recommend  that  the  HTT-1S  be  set  with  one  failure  at  a 
time to ensure that your testers are functioning correctly.   
 
 

A proper test plan would be constituted of the following tests: 

1.

 

Connect HTT-1S to the tester outputs. 

2.

 

Set all three switches on the HTT-1S to ‘PASS’. 

3.

 

Verify passing result on the Tester. 

4.

 

Set the Ground switch on the HTT-1S to ‘FAIL’; set both other switches to 

‘PASS’. 
5.

 

Verify Ground failure on the Tester. 

6.

 

Set the Leakage switch on the HTT-1S to ‘FAIL’; set both other switches to 

‘PASS’. 
7.

 

Verify Leakage current failure on the Tester. 

8.

 

Set the Hipot switch on the HTT-1S to ‘FAIL’; set both other switches to 

‘PASS’. 
9.

 

Confirm high voltage output by watching for a spark in the Arc Detect window 

of the HTT-1S; verify dielectric breakdown failure on the Tester. 
 

The  above  plan  is  presented  in  more  detail  in  the  Operation  Section  of  this 

Manual;  showing  proper  connection  of  the  HTT-1S  to  any  dielectric  tester  and 
any high current ground continuity tester you may wish to test. 
 

If any of the tests above return unanticipated results, you may have a problem 

with  your  tester.    It  is  recommended  that  it  be  removed  from  service  until  the 

Summary of Contents for HTT-1S

Page 1: ......

Page 2: ......

Page 3: ...tions Compliance West USA is proud to present you with your HTT 1S The operating instructions are included in the Manual so please retain this information for future use Thank you for your trust and c...

Page 4: ...ii...

Page 5: ...t Panel Features 7 Figure 3 HTT 1S OFFSET switch location 7 TTable 2 HTT 1S front panel Controls Indicators and Connectors 8 Setup Procedure Hipot Tester 9 Setup Procedure Ground Bond Tester 11 Figure...

Page 6: ...1...

Page 7: ...at a time to ensure that your testers are functioning correctly A proper test plan would be constituted of the following tests 1 Connect HTT 1S to the tester outputs 2 Set all three switches on the HT...

Page 8: ...3 problem can be resolved Section 4 of this manual also has troubleshooting information to help you determine if the HTT 1S is functioning correctly...

Page 9: ...ecifications of the HTT 1S Maximum current capability of GROUND test circuit 30 Amps Ground Test time should be 3 seconds Max and rest 10 s for currents above 10A Setting of PASS for Ground Circuit 0...

Page 10: ...5 Figure 1 Electrical Diagram of the HTT 1S...

Page 11: ...HTT 1S Function Tester This manual Return cable Output cable Remove the HTT 1S from the container and place it on a non conductive test bench Plug the cables follow instructions on Set up procedure To...

Page 12: ...7 Figure 2 HTT 1S Front Panel Features Figure 3 HTT 1S OFFSET switch location...

Page 13: ...e tester should stop testing and indicate a failure When set to PASS a normal passing result is simulated 5 Leakage Test Switch When set to FAIL an excess leakage condition is simulated and the tester...

Page 14: ...t fit the hipot ground bond tester For custom cables please contact Compliance West USA 1 Connect the hipot tester to the HTT 1S i On your hipot tester identify the high voltage output receptacle Conn...

Page 15: ...d GROUND OPEN test by setting the GROUND switch on the HTT 1S to FAIL All other switches on the HTT 1S should be set to PASS Begin testing by starting your hipot tester i Your hipot tester should indi...

Page 16: ...ound bond tester has an output current above 15A make sure the test time is not higher than 3 seconds to repeat another test wait for 10 seconds Resistors inside may overheat this can cause permanent...

Page 17: ...IL stop testing and sound a buzzer or otherwise warn the operator of a problem Unanticipated Results The HTT 1S test may cause your tester to give results not consistent with those anticipated by the...

Page 18: ...easured with a high precision resistance meter Switch designation Switch position Operating mode GROUND LEAKAGE HIPOT PASS PASS PASS No failure the tester connected to the HTT 1S should indicate test...

Page 19: ...14 Table 4 HTT 1S Resistance for ground circuit Ground switch position Resistance between Chassis and Return with 0 03 tolerances Pass 0 025 0 055 Fail 0 1 0 13...

Page 20: ...ce West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Reviews: