10 

 

2)

 

Ground OFFSET

 instructions:  

a)

 

With  the  HTT-1S  connected  to  your  Hipot  tester  (S  series  only!),  press 
the OFFSET switch and keep it pressed during all the subsequent actions.  

b)

 

 Press and hold the Ground Offset switch on the back of the Hipot Tester 
for  longer  than  5  seconds  until  the  display  shows  a  resistance  value  that 
equals to the offset.  

c)

 

To  verify  that  it  worked,  after  the  offset  is  done,  with  the  HTT-1S  still 
connected press the Ground Offset switch on the Hipot Tester for about 2 
to 3 seconds and a resistance value for ground will display, check 

Table 4

 

on this manual for the correct value for the ground resistance. 

3)

 

Conduct a simulated PASSING test by setting all three switches of the HTT-1S 

to the PASS (green) setting and initiating a hipot test by starting your hipot tester. 

i)

 

Your hipot tester should indicate ‘PASS’. 

 
4)

 

If  your  tester  checks  for  ground  continuity,  conduct  a  simulated  GROUND 

OPEN test by setting the GROUND switch on the HTT-1S to ‘FAIL’.  (All other 
switches on the HTT-1S should be set to ‘PASS’.)  Begin testing by starting your 
hipot tester. 

i)

 

Your hipot tester should indicate ‘FAIL’, stop testing, and sound a buzzer 
or otherwise warn the operator of a problem. 

 
5)

 

Conduct  a  simulated  EXCESS  LEAKAGE  test  by  setting  the  LEAKAGE 

switch on the HTT-1S to ‘FAIL’.  (All other switches on the HTT-1S should be 
set to ‘PASS’).  Begin testing by starting your hipot tester. 

i)

 

Your hipot tester should indicate ‘FAIL’, stop testing, and sound a buzzer 
or otherwise warn the operator of a problem. 

 
6)

 

Conduct a simulated HIPOT FAILURE test by setting the HIPOT switch on the 

HTT-1S to ‘FAIL’.  (All other switches on the HTT-1S should be set to ‘PASS’.)  
Begin testing by starting your hipot tester. 

i)

 

The  dielectric  breakdown  will  be  visible  as  an  arc  in  the  ARC  VIEW 
window of the HTT-1S. 

ii)

 

Your hipot tester should indicate ‘FAIL’, stop testing and sound a buzzer 
or otherwise warn the operator of a problem.  (If the current of the spark 
is above the leakage limit, the Excess Leakage failure may occur as well) 

 

 

Summary of Contents for HTT-1S

Page 1: ......

Page 2: ......

Page 3: ...tions Compliance West USA is proud to present you with your HTT 1S The operating instructions are included in the Manual so please retain this information for future use Thank you for your trust and c...

Page 4: ...ii...

Page 5: ...t Panel Features 7 Figure 3 HTT 1S OFFSET switch location 7 TTable 2 HTT 1S front panel Controls Indicators and Connectors 8 Setup Procedure Hipot Tester 9 Setup Procedure Ground Bond Tester 11 Figure...

Page 6: ...1...

Page 7: ...at a time to ensure that your testers are functioning correctly A proper test plan would be constituted of the following tests 1 Connect HTT 1S to the tester outputs 2 Set all three switches on the HT...

Page 8: ...3 problem can be resolved Section 4 of this manual also has troubleshooting information to help you determine if the HTT 1S is functioning correctly...

Page 9: ...ecifications of the HTT 1S Maximum current capability of GROUND test circuit 30 Amps Ground Test time should be 3 seconds Max and rest 10 s for currents above 10A Setting of PASS for Ground Circuit 0...

Page 10: ...5 Figure 1 Electrical Diagram of the HTT 1S...

Page 11: ...HTT 1S Function Tester This manual Return cable Output cable Remove the HTT 1S from the container and place it on a non conductive test bench Plug the cables follow instructions on Set up procedure To...

Page 12: ...7 Figure 2 HTT 1S Front Panel Features Figure 3 HTT 1S OFFSET switch location...

Page 13: ...e tester should stop testing and indicate a failure When set to PASS a normal passing result is simulated 5 Leakage Test Switch When set to FAIL an excess leakage condition is simulated and the tester...

Page 14: ...t fit the hipot ground bond tester For custom cables please contact Compliance West USA 1 Connect the hipot tester to the HTT 1S i On your hipot tester identify the high voltage output receptacle Conn...

Page 15: ...d GROUND OPEN test by setting the GROUND switch on the HTT 1S to FAIL All other switches on the HTT 1S should be set to PASS Begin testing by starting your hipot tester i Your hipot tester should indi...

Page 16: ...ound bond tester has an output current above 15A make sure the test time is not higher than 3 seconds to repeat another test wait for 10 seconds Resistors inside may overheat this can cause permanent...

Page 17: ...IL stop testing and sound a buzzer or otherwise warn the operator of a problem Unanticipated Results The HTT 1S test may cause your tester to give results not consistent with those anticipated by the...

Page 18: ...easured with a high precision resistance meter Switch designation Switch position Operating mode GROUND LEAKAGE HIPOT PASS PASS PASS No failure the tester connected to the HTT 1S should indicate test...

Page 19: ...14 Table 4 HTT 1S Resistance for ground circuit Ground switch position Resistance between Chassis and Return with 0 03 tolerances Pass 0 025 0 055 Fail 0 1 0 13...

Page 20: ...ce West USA is available Phone 800 748 6224 Hours 8 00 AM 4 00 PM Pacific Time Also available on our web site at www compwest com Contact Compliance West USA 650 Gateway Center Way Suite D San Diego C...

Reviews: