Setup Procedure: Ground Bond Tester
This procedure will describe connection of the HTT-1S to a ground bond
tester. Because of the varying types of tester outputs, only return banana to
banana cable is provided with the HTT-1S that can be used on a Ground bond
tester. For custom cables, please contact Compliance West USA.
If the ground bond tester has an output current above 15A, make sure
the test time is not higher than 3 seconds, to repeat another test wait for 10
seconds. Resistors inside may overheat, this can cause permanent damage
and miss readings.
1)
Connect the ground bond tester to the HTT-1S.
a)
On your ground bond tester, plug the return lead of the high current
output to the black return banana receptacle of the HTT-1S.
b)
On your ground continuity tester, plug the HTT-1S into the high current
receptacle.
Figure 5: Connection to a Ground bond tester