By default, all of the channels are active at the beginning of the session. To deselect a channel, 

simply press the corresponding button.

Notes:

 depending on the program, the diagram on the middle of the screen can change. There are 

two types of diagram listed below.

Refer to above screenshot, In this case only channel 1 is active. Any change of intensity will only 

be performed on channel 1.

INCREASE THE STIMULATION INTENSITIES ON THE SELECTED CHANNELS.

1. WORK PHASE/RELAXATION PROGRAM

These programs always begin with a warm-up phase. After this warm-up phase, a work cycles 

phase followed by relaxation occurs (the number of cycles depends on the program) and when all 

cycles are completed, the program ends with a recovery phase.

2. WARM UP, RECOVERY TYPE PROGRAM

These program consist of a single phase and have no work phase/relaxation cycle.            

These are warm up, recovery, or even pain type program. During this type of program, 

frequency variations can occur.

A. 

Warm up phase  

B. 

Work phase   

C. 

Relaxation phase  

D. 

Recovery phase    

E. 

Work time phase which consists of work 

phase and relaxation phase   

A. 

Work time phase

By pressing on the central confirmation button during stimulation, the device goes into pause. At 

this point it is possible to skip the current phase or exit the program.
N.B.: The session restarts with intensities equal to 75% of those used prior to the interruption.

At the end of the session a screen with a check 

mark appears. Pressing on the central confirmation 

or right buttons return you to the home page. To 

turn off the device, hold down the On/Off button 

for 5 seconds.

F: CONTINUE A STIMULATION PROGRAM

G: END A STIMULATION PROGRAM

B

C

A

A. 

skip the current phase.      or 

exit the program 

B. 

Resume the stimulation 

session

C. 

Skip the current phase.      or 

exit the program

Summary of Contents for WIRED 3.0

Page 1: ...WIRED 3 0 Muscle Stimulators Edge 3 0 Performance 3 0 Sport Elite 3 0 READ THE MANUAL BEFORE OPERATING THIS PRODUCT...

Page 2: ...ody parts required for normal function Critical ischemia of lower limb Seriously compromised oxygen carrying blood flow to the legs feet and toes lower limb that may result in loss of the limb if left...

Page 3: ...contraindication indicates a condition or situation in which a device should not be used NOTE DANGEROUS VOLTAGE Text with a Dangerous Voltage indicator serves to inform the user of possible hazards re...

Page 4: ...lation session on a person who is standing The first five minutes of stimulation must always be performed on a person who is sitting or lying down In rare instances people of a nervous disposition may...

Page 5: ...ive medium or alternate electrode placement If you have suspected or diagnosed heart disease you should follow precautions recommended by your physician If you have suspected or diagnosed epilepsy you...

Page 6: ...imulation accurately reproduces the process observed during a voluntary contraction The stimulator sends an electrical current impulse to the nerve fibers exciting them This excitation is then transmi...

Page 7: ...referred to as gates These gates determine which impulses are allowed to continue to the brain The gates prevent the brain from receiving too much information too quickly Since the same nerve cannot...

Page 8: ...cm 12 9mA in Always turn off the stimulator before moving or removing any electrodes during a session Do not place the electrodes in water Do not apply solvents of any kind to the electrodes For best...

Page 9: ...ecome directly accessible after turning on the device NOTE Programs vary depending on device The programs done will automatically be placed in the favorites menu The favorites menu can contain up to 1...

Page 10: ...r selection A Select the desired level B Confirm your selection C Return to the previous step Note if the level option in setting is turned off this step would be canceled D HOW TO ACCESS SETTINGS The...

Page 11: ...eans the corresponding channel is selected After user choose the desired channel press the central confirmation button to confirm your selections then follow the tips on screen press up to start to pr...

Page 12: ...phase relaxation cycle These are warm up recovery or even pain type program During this type of program frequency variations can occur A Warm up phase B Work phase C Relaxation phase D Recovery phase...

Page 13: ...y is completely discharged It is no longer possible to use the device Recharge it immediately The charge state of the battery is indicated by a small battery icon on the top right screen A BATTERY LEV...

Page 14: ...pe of effort are for example the 400 and 800 meters one kilometer cycle races and 100 meters swimming It is appropriate for many other sports based on duration such as cycling which makes repeated dem...

Page 15: ...by placing the electrode connected to the positive pole red connection strategically When working with a muscle stimulation program involving muscle contractions it is important to place the positive...

Page 16: ...th others When removing your pads from your skin peel from the side of the pad Do not pull the pad off by the wire When you store your electrode pads after your session make sure to stick them on the...

Page 17: ...ls REACH Caution This is an indication for protection against ingress of water and particulate matter The mark IP22 on your unit means your unit is protected against solid foreign objects of 12 5mm di...

Page 18: ...11 RF Emissions Group 1 The Compex Wired 3 0 uses RF energy only for its internal operation Consequently its RF emissions are very low and are unlikely to interfere with any adjacent electrical device...

Page 19: ...ngth in the location in which the Compex Wired 3 0 are used exceeds the applicable RF compliance level above the Compex Wired 3 0 should be observed to verify normal operation If abnormal performance...

Page 20: ...ilure in the Product caused by product misuse including but not limited to the failure to provide reasonable and required maintenance or any use that is inconsistent with the Product User Manual COMPA...

Page 21: ...40 41 CONTACTINFORMATION Distributed By DJO LLC 1430 Decision Street Vista CA 92081 1 877 266 7398 877 COMPEX8 JKH Health Co LTD Shajing Baoan District Shenzhen China Made in China...

Page 22: ...t 2020 by DJO LLC Individual results may vary Neither DJO LLC nor any of its subsidiaries dispense medical advice The contents of this document do not constitute medical advice Rather please consult y...

Reviews: