IE-SNC-T03(4)10+T5410-06
Instruction Manual
Device type
Temperature Pressure Output
Galvanic isolated output
T0310
✓
-
RS232
-
T0410
✓
-
RS485
✓
T5410
✓
✓
RS485
✓
Device is designed for measurement of ambient temperature at °C or °F, device T5410
support atmospheric pressure measurement at hPa, PSI, inHg, mBar, oz/in
2
, mmHg, inH
2
O and kPa
too. It is possible to set pressure correction for altitude (offset). Device setting is provided by user’s
software, see latter. Device T0310 communicate via link RS232, devices T0410 and T5410 via
link RS485. Supported communication protocols are Modbus RTU, protocol compatible with
standard Advantech-ADAM, ARION and communication with HWg–Poseidon devices. Devices
are preset from manufacturer to Modbus RTU communication protocol. If you would like to use
different communication protocol then Modbus RTU, it is necessary preset them through user’s
software – see latter. Temperature and pressure sensors are non-removable device parts, pressure
sensor is placed inside the electronic case. Measured values are optionally shown on dual line LCD
display. Display can be switched OFF totally too.
Use user’s software TSensor for setting of all device parameters (recommended).
It is free to download at
. It supports make the adjustment of the device
too. This procedure is described at file „
Calibration manual.pdf
“ which is installed commonly with
the software. Change of some parameters is possible to do without user’s software with Windows
hyperterminal (change of communication protocol, its parameters, LCD display setting).
It is described in file “
Description of communication protocols of Txxxx series
” which is free
to download at the same address.
Transmitter version T
xxxx
L
with watertight male connector instead of a cable gland
(RS232) or glands (RS485) is designed for easy connection/disconnection of communication cable.
Male Lumberg connector RSFM4 has protection IP67.
Models marked T
xxxx
Z
are non-standard versions of the transmitters. Description is not
included in this manual.
Please read instruction manual before the first device connection.
Device setting from the manufacturer
If special setting was not required in the order device is set from the manufacturer to the following
parameters:
communication protocol:
Modbus RTU
device address:
01
communication speed:
9600Bd, without parity, 2 stop bits
display:
switched ON
temperature unit:
°C
pressure unit:
hPa
correction for altitude:
0 hPa
(absolute pressure)