background image

 

check free space in data memory, optionally erase data memory of the logger 

 

enter password if protection against unauthorized manipulation with the logger is necessary 

 

Logging interval between to subsequent measurements is specified by the user. Memorizing of the 
first value is synchronized with the internal real time clock, so the logging is performed at sharp 
multiples of minutes, hours and days. E.g. after starting the logging with the 15 minute interval the 
first value is not stored immediately, but after the internal clock gets the status of a quarter, a half or 
a whole hour. After starting the logging with the 6 hours interval the first value is stored at that 
whole hour to perform the storing also at 00.00, i.e. at the beginning of the day. First storing is 
performed at 6.00,12.00, 18.00 or 00.00hour - at the hour from the above nearest to the logging 
start. After communication with computer or after start by magnet logger automatically waits for 
the nearest whole multiple of time and then first measurement is performed. This is also necessary 
to take into account when setting time of automatic logger switch ON. 

Notice:

 If logger operates as permanently connected to the computer, using of magnet 

start/stop is disabled. 

 To enable the logger control by magnet is suitable only in cases, when possibility of 

unauthorized manipulation to the logger operation is eliminated. 

 

If logger is controled by external signal (i.e. by binary state at input 4), logger will be switched 

OFF (when programmed binary state at input 4 appears) with the delay up to 10 s! 

User description of binary states is displayed only at the data list on the computer. On the 

logger LCD indication ON and OFF is displayed permanently. 

Selection of „display of increments“ at the card of counter channel setting causes record of 

counter increments in time specified by logging interval. On the logger LCD counter values are 
displayed (not increments). 

 

Logger connection to monitored devices 

Connect logger to monitored device by shielded cable with 9 pins female CANON connector. 

If IP67 protection is required, it is necessary to use watertight connector (optional accessory). 
Channel 1 of the logger (IN1) works as counter at pin 6 of the connector, channel 4 works as binary 
input at pin 1 of the connector. Both inputs have common ground (GND) at connector pins 5 a 9. 
Those pins serve for connection of the cable shielding. As monitored signal use quality mechanical 
contact (reed contact) or device with „open collector“ or device with binary voltage output. Due to 
high input impedance it is necessary to avoid presence of increased electromagnetic interference 
(e.g. electric switchboards, motors, device containing power current switching elements - 
contactors, relays or pulse converters etc.). 

If logger will be controlled (switch ON and OFF) by external binary signal, connect output of 

the control device to pin 1 (input 4 of the binary channel) and pin 5 (signal ground). It is necessary 
to allow selection „Logger control by external signal“. At the same time set required binary level 
for logger switch ON at the card of channel 4. In switching ON by external signal logger reacts 
immediately, in switching OFF reacts with delay of maximum 10 s. 

Notice:

 pin 9 is galvanic connected to pin 5 (both GND). If other signal source is used than 

potential-less mechanical contact, note, that signal grounds of both channels will be connected 
together – grounds of both devices must be mutually galvanic isolated or must be at the same 
potential! 

IE-LOG-S7021-04 

 

  5 

Summary of Contents for S7021

Page 1: ...LOGGER S7021 Dual channel logger with counter and binary inputs Instruction for use...

Page 2: ...aximum value and logger is switched OFF with the message on the LCD If needed counter mode can be selected when after fulfilling counter overflows and continues from value 0 Logging mode can be adjust...

Page 3: ...me clock adjustable from computer integrated calendar including leap years Error of internal RTC 200 ppm i e 0 02 17 28 s in 24 h Power Lithium battery 3 6 V size AA Battery life typical data download...

Page 4: ...setting of the instrument accordingly to the user needs menu Configuration Setting of instrument parameters Before logging start it is necessary check or optionally set the logger real time clock sele...

Page 5: ...e logger LCD indication ON and OFF is displayed permanently Selection of display of increments at the card of counter channel setting causes record of counter increments in time specified by logging i...

Page 6: ...mperature range battery life can be reduced to 75 i e if displayed remaining battery capacity drops below 25 it is recommended to replace the battery If LCD is switched ON actual readings are displaye...

Page 7: ...ation decimal point goes out logger would be switched OFF Display goes out automatically after 30 s Remove magnet from slots anytime during display is ON or later Activation of switched OFF LCD for 30...

Page 8: ...clock setting below critical limit or battery disconnection for longer period than approximately 30 s after display switch ON during date and time display all four arrows appears as warning to check o...

Page 9: ...s identical with the above procedure for switching ON If decimal point does not appear after 1 s it is necessary remove the magnet and repeat the procedure Logger connection to monitored device by the...

Page 10: ...l logger settings remain unchanged In the opposite case check by means of PC program all settings especially the real time clock in logger Attention inserted battery with incorrect polarity causes log...

Reviews: