Battery replacement
Low battery is indicated on the display by blinking of reading "BAT". It can be displayed
permanently, if battery voltage is too low. Replace the battery for new one. If logger is operated
often in temperature below -5°C or over +35°C and PC program indicates remaining battery
capacity below 25% it is also recommended to replace the battery. Applied is Lithium battery
3.6 V, size AA. Battery is located under logger lid.
Warning:
near battery fragile glass reed contact is located – be careful not to damage it. Be
careful in battery replacement!
Replacement procedure:
•
switch off the logger by the PC program or by magnet (if low battery allows)
•
unscrew four corner screws and remove the lid
•
remove old battery by pulling the glued tape
•
insert new battery
respecting the correct polarity
(see s and – near battery holder). If
you connect new battery up to 30 s, all logger settings remain unchanged. In the opposite case
check by means of PC program all settings, especially the real time clock in logger.
Attention,
inserted battery with incorrect polarity causes logger damage!
•
put the lid back again and screw four screws. Be careful to have the rubber sealing properly in
the slot and tighten the screws properly to ensure the water resistance of the instrument.
•
connect logger to the computer and
write to it the information on battery replacement
(menu
Configuration/Battery replacement). This step is necessary to evaluate properly the remaining
battery capacity.
Old battery or logger itself (after its life) is necessary to liquidate ecologically!
Instruments passed the following electromagnetic compatibility (EMC) tests:
radiation: EN 55022
class B
immunity: EN 61000-4-2
(levels 4/8 kV, class A)
EN 61000-4-3
(intensity of electromagnetic field 3 V/m, class A)
EN 61000-4-4
only types with external probes (levels 1/0,5 kV, class A)
EN 61000-4-6
only types with external probes (intensity of electromagnetic
field 3 V/m, class A)