SNMP
Using SNMP protocol you can read actual measured values, alarm status and alarm
parameters. Via SNMP protocol is also possible to get last 1000 measured values from
history table. Writing via SNMP protocol is not supported. It is supported
SNMPv1
protocol version only. SNMP used
UDP port 161
. OID keys description can be found in
the MIB table, which can be obtained from device website or from your distributor. The
password for reading is factory set to
public
. The changes can be made using TSensor
software or web interface. OID keys:
OID
Description
Type
.1.3.6.1.4.1.22626.1.5.1
Device informations
.1.3.6.1.4.1.22626.1.5.1.1
Device name
String
.1.3.6.1.4.1.22626.1.5.1.2
Serial number
String
.1.3.6.1.4.1.22626.1.5.1.3
Device type
Integer
.1.3.6.1.4.1.22626.1.5.2.ch
Measured value (where ch is channel number)
.1.3.6.1.4.1.22626.1.5.2.ch.1
Channel name
String
.1.3.6.1.4.1.22626.1.5.2.ch.2
Actual value – text
String
.1.3.6.1.4.1.22626.1.5.2.ch.3
Actual value
Int*10
.1.3.6.1.4.1.22626.1.5.2.ch.4
Alarm on channel (0/1/2)
Integer
.1.3.6.1.4.1.22626.1.5.2.ch.5
High limit
Int*10
.1.3.6.1.4.1.22626.1.5.2.ch.6
Low limit
Int*10
.1.3.6.1.4.1.22626.1.5.2.ch.7
Hysteresis
Int*10
.1.3.6.1.4.1.22626.1.5.2.ch.8
Delay
Integer
.1.3.6.1.4.1.22626.1.5.2.ch.9
Unit
String
.1.3.6.1.4.1.22626.1.5.2.ch.10
Alarm on channel – text
String
.1.3.6.1.4.1.22626.1.5.3.1.0
SNMP Trap text
String
.1.3.6.1.4.1.22626.1.5.4.1.1.ch.nr
History table value
String
When alarm occurred a warning messages (trap) can be sent to selected IP addresses.
Addresses can be set using TSensor software or web interface. Traps are sent via
UDP
protocol on
port 162
. The device can send following traps:
Trap
Description
0/0
Reset of the device
1/0
Settings was changed
6/0
Testing Trap
6/1
NTP synchronization error
6/2
E-mail sending
error
SMTP server login error
6/3
SMTP authentication error
6/4
Some error occurred during SMTP communication
6/5
TCP connection to server cannot be opened
6/6
SOAP message
sending error
SOAP file not found inside web memory
6/7
MAC address can't be obtained from IP address
6/8
TCP connection to server cannot be opened
6/9
Wrong response code from the SOAP server
6/11 – 6/15
Upper alarm on channel
6/21 – 6/25
Lower alarm on channel
6/31 – 6/35
Clearing alarm on channel
6/41 – 6/45
Measuring error