Page 24 of 30
INSTRUCTION MANUAL
SEC TI ON 5
-
CALIB RATION & TE ST PROCED URE S - MI L-STD-461G (CS114)
Rev081219
4) Monitor the EUT for degradation of performance; and, whenever susceptibility
is noted, determine the threshold level in accordance with Section 4.3.10.4.3 of
MIL-STD-461G.
5) Set the frequency of the signal generator and measurement equipment to the
next test frequency.
6) Repeat Steps 3-5 until the test has been performed at all test frequencies.
7) Repeat Steps 2-6 until all required EUT cables have been tested.