Page 27 of 30
INSTRUCTION MANUAL
SEC TI ON 6
-
CALIB RATION & TE ST PROCED URE S
–
RTCA D O-160 (Secti on 20 )
Rev081219
1) Set up the equipment as described in the applicable standard or approved
test plan. An example setup diagram is given in Figure 18.
2) Turn on all equipment and allow sufficient time for stabilization. Set the
frequency of the signal generator and measurement equipment to 10 kHz,
unmodulated.
3) Slowly increase the amplitude of the signal generator until the value indicated
on Spectrum Analyzer/EMI Receiver #1 reaches the appropriate amplitude
(see section 6.2).
4) Record the present frequency and the measured forward power value
indicated on Spectrum Analyzer/EMI Receiver #2.
5) Set the frequency of the signal generator and measurement equipment to the
next test frequency (see section 6.1).
6) Repeat Steps 3-5 until all test frequencies have been calibrated.
NOTE:
Typical DO-160, Section 20 Forward Power Levels for the CLCI-400 are shown in
Figure 9, in Section 3.7 of this document.
Summary of Contents for CLCI-400
Page 2: ...Page 2 of 30 INSTRUCTION MANUAL TABLES OF CONTENTS Rev081219...
Page 3: ...Page 3 of 30 INSTRUCTION MANUAL TABLES OF CONTENTS Rev081219...
Page 4: ...Page 4 of 30 INSTRUCTION MANUAL TABLES OF CONTENTS Rev081219...
Page 6: ...Page 6 of 30 INSTRUCTION MANUAL SECTION 2 PRODUCTS AVAILABLE FROM COM POWER Rev081219...
Page 12: ...Page 12 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...
Page 13: ...Page 13 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...
Page 14: ...Page 14 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...
Page 15: ...Page 15 of 30 INSTRUCTION MANUAL SECTION 3 PRODUCT INFORMATION Rev081219...