PULSAR
®
PLUS 55-S SUCTION BLAST CABINET
Page 20
© 2021 CLEMCO INDUSTRIES CORP.
www.clemcoindustries.com
Manual No. 30423
5. When "A" appears on the display, do not press
anything; after a few seconds, the unit resumes
operation with the "A" function.
5.9.5 Reset timer subranges to factory settings
NOTE: The OFF subrange is factory set for seconds, as
shown in Figure 25. The ON subrange is factory set for
milliseconds. Each value shown on the LED is 10
milliseconds (1/100 of a second), as shown in Figure 26.
There is no need to change subranges; if subranges are
accidentally changed, reset them to factory settings, as
follows:
5.9.5.1
Reset OFF (time between each pulse) subrange
to factory setting of seconds, as shown in Figure 25:
1. Press the left "off/arrow down" button and OF
appears briefly on the display.
2. If the min (minute) subrange light is lit, decrease OFF
subrange by pressing the left "off/arrow down" button
until the sec (seconds) light is lit.
3. If the min (minute) and sec (seconds) subrange lights
are both lit, increase OFF subrange by pressing the right
"on/arrow up" button until only the sec light remains lit.
4. Set OFF time to suggested time, per Section 5.9.2.
TEST
of
f
on
on
FACTORY SETTING FOR
OFF
SUBRANGE
Figure 25
5.9.5.2
Reset ON (duration of pulse) subrange to factory
setting of milliseconds, as shown in Figure 26.
1. Press the right "on/arrow up" button, and ON appears
briefly on the display.
2. To decrease ON subrange, press the left "off/arrow
down" button until the min and sec indicator lights are
both lit, indicating that the subrange is set for
milliseconds x 10. For example, if the seconds subrange
light is lit, press the "off/arrow down" button past 0
seconds to lower the subrange to milliseconds.
3. Set ON time to 15 (15 milliseconds x 10 = 150
milliseconds), per Section 5.9.3.
TEST
of
f
on
on
FACTORY SETTING FOR
ON
SUBRANGE
Figure 26
5.10
Using the Optional Manometer to Set
Reclaimer Static Pressure
NOTE: These instructions and the illustrations in Figure
27 show two methods for taking static-pressure readings
(negative pressure) on Pulsar reclaimers. Use the
method best suited for the application. The instructions
explain the processes for taking occasional readings and
show how to install a fitting in the reclaimer body for
taking frequent readings. Taking readings at different
locations could produce slightly different readings. The
readings are reference points, so readings should be
taken using the same method each time they are taken.
5.10.1
Refer to instructions packed with the manometer
for preparing and operating the manometer.
5.10.2
Open both manometer valves (elbows), at the
top of the manometer, per the manometer instructions.
5.10.3
Push one end of the 3/16" ID tubing provided
with the manometer, onto either one of the two valves.
5.10.4
Attach the manometer at a convenient location,
as shown in the example in Figure 27. Magnets on the
manometer hold it onto any metal surface on the power
module. The manometer must be vertical, so the fluid is
level on both sides.
5.10.5
Adjust the slide rule to align the zero with the
fluid level. Refer to Figure 28.
The sec (seconds)
indicator light is lit,
indicating the subrange is
set for seconds.
OF
OFF Subrange
1 2
3
1
2
ON
The min (minute) and sec
(seconds) indicator lights
are lit, indicating the
subrange is set for
milliseconds x 10.
ON Subrange