![CleanSpace CS302 User Instructions Download Page 28](http://html1.mh-extra.com/html/cleanspace/cs302/cs302_user-instructions_2620497028.webp)
28
you leave the contaminated area may help).
6.
Leave the contaminated area immediately if any of the following conditions occur:
a.
Any part of the system is damaged.
b.
Airflow into the mask decreases or stops.
c.
The battery or filter alarms are triggered (even if only the audible or only the visible
alarm triggers).
d.
Breathing becomes difficult.
e.
You feel dizzy or your vision is impaired.
f.
You smell or taste contaminants.
g.
Your face, eyes, nose or mouth become irritated.
h.
You suspect the concentration of contaminants may have reached levels at which this
respirator may no longer provide adequate protection.
7.
Follow the exiting and decontamination procedures as documented in the workplace Respiratory
Protection Program.
To remove the respirator, find the Mask Release Pincer and pinch to release the mask from the blower. See Figure
25.
Figure 25 – Mask Release Pincers
Note: The respirator motor will automatically switch off and go into Standby Mode in approximately 10 seconds
when no breath is detected. In Standby Mode, if after three (3) minutes no breath is detected, the Respirator will
automatically go into Off Mode. In Off Mode, all the battery indicator lights will be off and the motor will not be
triggered by the wearer’s breathing.
WARNING
DO NOT REMOVE THE RESPIRATOR UNTIL YOU HAVE
VACATED THE CONTAMINATED AREA unless you have pressing health reasons to do so (for
instance you are experiencing dizziness and believe removing the respirator while you leave the
contaminated area may help).
17. Working in Noisy Environments
In most noisy environments the CleanSpace HALO alarms are still clearly audible to the wearer. But if there is a
chance that you will not hear the alarms, you must take the following extra precautions.
Never enter the contaminated zone unless all three green battery indicator lights are illuminated.
Never work in the high noise area for more than four (4) hours. At the end of four hours, leave the
contaminated zone and check that
o
the filter blocked light is not illuminated; AND
o
all three green battery indicator lights are still illuminated.
If two or fewer green battery indicator lights are illuminated, recharge CleanSpace HALO until 3 lights are
showing again (and the 3
rd
one has stopped flashing).
If the filter blocked light is illuminated, change the filter.
Be particularly aware of difficulty breathing or of the air flow stopping. If these things happen, exit the
contaminated zone immediately.
Summary of Contents for CS302
Page 8: ...8 Figure 1 System Components 9 System Components...
Page 46: ...46 2 1 LED CleanSpace HALO 3 1 Steri Plus HALO BIO CS3039 CS3027 IPA...
Page 55: ...55 PAF 0025 PortaCount S005 7174 PortaCount 22 PAPR CleanSpace HALO OSHA 1901 134 NIOSHI...
Page 57: ...57 13 CleanSpace HALO 5 1 HEPA HEPA 18 HEPA CleanSpace HALO 3 LED 19 2 20 C 1 2...
Page 59: ...59 CleanSpace CleanSpace HALO 3 5 5 8 6 9 4...
Page 60: ...60 7 2 FRONT 2 9...
Page 61: ...61 2 HALO OSHA 1910 134 Steri Plus Steri Plus Steri Plus 11 11 A B 12 12 10...
Page 62: ...62 13 13 14 15 16 17 18 19 1...
Page 63: ...63 20 CleanSpace HALO 5 HALO BIO CS3024 HALO BIO 14 CleanSpace HALO 21...
Page 64: ...64 PAF 1009 22 23 2 6 STERI PLUS PAF 1111 22 23...
Page 66: ...66 HALO BIO CS3029 HALO BIO HALO BIO 2 2 2 HALO BIO HALO BIO HALO BIO HALO BIO...
Page 67: ...67 15 CleanSpace HALO 6 1 HEPA HEPA 18 CleanSpace HALO 3 LED 19...
Page 69: ...69 3 LED 19 4 HEPA CleanSpace HALO 5 CleanSpace Technology...
Page 70: ...70 5 10 6 CleanSpace PAPR 4...
Page 71: ...71 2 1...
Page 72: ...72 CleanSpace CleanSpace 16 1 2 3 17 4 5 6 a b c d e f...
Page 73: ...73 g h 7 9 28 28 10 3 17 CleanSpace HALO 3 4 4 3 1 2 3 3 CleanSpace HALO 18...
Page 74: ...74 CleanSpace 1 2 2 1 LED 1 15 CS3027 HALO BIO CS3039 Steri Plus...
Page 82: ...82 2 122 F 50 C 2...
Page 83: ...83 HEPA CleanSpace HALO CleanSpace 21 CleanSpace HALO 36 F 20 C 13 22...
Page 84: ...84 PAF 0025 PortaCount S005 7174 PortaCount 1 TSI PortaCount CleanSpace TSI PortaCount...
Page 85: ...85 CleanSpace 1000 1000 1000 CleanSpace 25 23 HALO HALO BIO...