5
English
1
2
3
4
5
6
7
• Replace the battery immediately if the battery warning symbols appears on
the display. A weak battery can cause incorrect results and thereby pose a safety
hazard.
• Neither the meter nor its accessories may be dismantled or modified in any way.
• Remove the battery from the meter if it is not to be used for an extended period.
• Turn the meter off after use.
• If the meter is placed close to a strong magnetic field, it might cause incorrect
readings which can be corrected by removing the cause of the interference.
Buttons and functions
1. Display
2. [ HOLD ] Press to lock
the measured value.
3.
: connection for red
test lead when measuring DC
current up to 10 A.
4.
: connection for
red test lead when measuring
voltage, DC current up to
0.2 A and resistance, and for
testing diodes and continuity.
5. [
]
Switch display lighting
on/off.
Note:
The display
light does not switch off
automatically, it needs to be
switched off manually.
6. Function selector switch
7.
: connection for
the black test lead (common).
Display symbols
: The batteries needs replacing.
:
The test leads are connected with the wrong polarity to either the instrument or
the object being measured.
OL: The reading is outside of the measuring range.