ASTRA® VR Endovaginal/Endorectal Probe Reprocessor
Installation Manual
M0065_C
7
ASTRA® VR
•
Water Supply Connection:
The device is designed to accommodate a cold potable water supply. The installer shall cut
the supplied water inlet hose to the desired length and add the supplied 3/8” male hose barb
(A6041-01) and hose clamp (A6132-01). Install the 3/8” NPT Valve Coupling (A6166-01) into
the 3/8” NPT (F) shut-off valve not supplied. A water pressure regulator (not supplied) is
required if inlet water pressure exceeds water supply specifications of 50 psi (see section 3).
o
Connect terminated end of tubing assembly to the water inlet port on the right
side of the device (see Figure 4). Ensure the connector fully locks into place.
o
Turn on water supply to unit and check for leaks.
•
Drain Connection:
The installer shall cut the drain hose to the desired length and terminate the open end of the
tubing assembly into a sanitary waste drain. This can be into a typical dishwasher type
tailpiece under a sink, a floor standpipe or directly into a waste sink (secure line to prevent
drain water from splashing). Ensure any required air gaps are maintained.
Figure 4 –
Drain and water quick-disconnect fittings
E.
Electrical Connection:
It is the installer’s responsibility to ensure the device is installed in a manner which is
compliant with all applicable electrical codes and facility requirements. Install a 15A Hospital
Grade GFCI receptacle within 8 ft. of the designated site of the installed device.
Plug in supplied hospital grade power cord into the right side of the unit and into a GFCI wall
receptacle. Switch on green illuminated power switch. If switch does not illuminate, make
sure GFCI receptacle has been reset.
CAUTION:
Use of a power cord not supplied by CIVCO Medical
Solutions may affect the safety and performance of this
device.
NOTE:
Do not attempt to use device – water and air filters have not yet been installed.
Once complete, device is ready for in-service. Turn off power. Fill out attached checklist
and return to CIVCO Medical Solutions.