ASTRA® VR Endovaginal/Endorectal Probe Reprocessor
Installation Manual
M0065_C
8
ASTRA® VR
6.
INSTALLATION CHECKLIST
Once the plumbing, electrical and mounting connections have been installed, a photograph
of the installation should be taken, a list of probe models to be used with the ASTRA VR
complied and the following completed checklist should be emailed to CIVCO Medical
Solutions.
:
Checked
Electrical
•
Hospital Grade GFI installed within 8 ft. of device location
□
•
Hospital Grade GFI electrically tested
□
•
Device has passed hospital’s biomed electrical leakage test
□
Plumbing
•
Potable water supply with fitting and stop valve within 8 ft.
□
o
Water supply pressure within 30-50 psi
□
o
Water supply flow rate greater than 1 gpm
□
o
Water supply temperature within 50 -80 F
□
•
Drain available within 8 ft.
□
Physical
•
Mounting bracket and probe holders installed
□
•
Minimum clearance requirements met
□
Environmental
•
Environmental requirements met
□
Site Photograph
•
Photograph attached of intended installation site
□
Other
•
All parts shipped with ASTRA VR accounted for (see section 4)
□
•
Bottles of HLD available for in-service
(circle one)
o
Metricide™ OPA
o
Resert® XL
o
Cidex® OPA
□
•
List of probe models to be used with ASTRA VR attached
□
Facility Name: ____________________________________________________________
Contact Person: _____________________ Email: _____________________________
Telephone: _________________________ Desired In-service date: _______________