ASTRA® VR Endovaginal/Endorectal Probe Reprocessor
Installation Manual
M0065_C
6
ASTRA® VR
B.
Mounting the Unit Wall Bracket:
The ASTRA VR requires clear wall space of 24” wide and 57” high. A wall bracket is shipped
with the unit to hold the unit securely against the wall for added safety. Refer to Figure 2 for
mounting location of the bracket.
Mark the position of the screw holes as per Figure 2 and install the wall bracket (A1053-01)
with the EZ Anchors (A6164-01) supplied as shown below:
Figure 3
Wall Bracket Installation
Using two people, remove the device from the packaging. Using the lift handles on the sides
of the unit, lift the unit slightly over the wall bracket and position into place. Ensure weight of
device is on feet and device is level using the adjustable feet.
CAUTION:
The device weight exceeds 50lbs (22.7kg). When lifting
loads heavier than 50 pounds, use two or more people
to lift the load.
C.
Mounting Probe Connector Holders
Two probe connector holders (A1064-01) are supplied with the device and should be
installed on the left side of the unit, refer to Figure 2 for mounting locations. Install probe
connector holders with the EZ Anchor kit (A6164-01) supplied. Once mounted to the wall,
attach probe connector holder pads; back (A9064-01) and base (A9065-01) in each holder.
D.
Plumbing Installation
It is the installer’s responsibility to ensure the device is installed in a manner which is
compliant with all applicable plumbing codes and facility requirements. Plumbing installation
requires a connection to a waste drain and potable cold water supply. Consult your facility
manager to determine the best installation method for your particular situation.
The device includes two tubing assemblies: one for each of the drain and water supply
circuits. These assemblies are terminated on one end only. The terminated end is for
installation to the right side of the device. The opposite ends are not terminated and the
installer determines the method of termination as indicated below.