background image

ASTRA® VR Endovaginal/Endorectal Probe Reprocessor  

Installation Manual 

M0065_C

 

ASTRA® VR 

 

B.

 

Mounting the Unit Wall Bracket:  

 

The ASTRA VR requires clear wall space of 24” wide and 57” high. A wall bracket is shipped 

with the unit to hold the unit securely against the wall for added safety. Refer to Figure 2 for 

mounting location of the bracket.  

 

Mark the position of the screw holes as per Figure 2 and install the wall bracket (A1053-01) 

with the EZ Anchors (A6164-01) supplied as shown below: 

 

 

Figure 3  

Wall Bracket Installation

 

 

Using two people, remove the device from the packaging. Using the lift handles on  the sides 

of the unit, lift the unit slightly over the wall bracket and position into place. Ensure weight of 

device is on feet and device is level using the adjustable feet. 

 

CAUTION:

     The device weight exceeds 50lbs (22.7kg). When lifting 

loads heavier than 50 pounds, use two or more people 

to lift the load. 

 

C.

 

Mounting Probe Connector Holders 

 

Two probe connector holders (A1064-01) are supplied with the device and should be 

installed on the left side of the unit, refer to Figure 2 for mounting locations. Install probe 

connector holders with the EZ Anchor kit (A6164-01) supplied. Once mounted to the wall, 

attach probe connector holder pads; back (A9064-01) and base (A9065-01) in each holder.  

 

D.

 

Plumbing Installation  

 

It is the installer’s responsibility to ensure the device is installed in a manner which is 

compliant with all applicable plumbing codes and facility requirements. Plumbing installation 

requires a connection to a waste drain and potable cold water supply. Consult your facility 

manager to determine the best installation method for your particular situation.  

 

The device includes two tubing assemblies: one for each of the drain and water supply 

circuits. These assemblies are terminated on one end only. The terminated end is for 

installation to the right side of the device. The opposite ends are not terminated and the 

installer determines the method of termination as indicated below. 

 

 

Summary of Contents for Astra VR

Page 1: ... registered trademark of CIVCO Medical Solutions All other trademarks are property of their respective owners Printed in the USA CIVCO Medical Solutions 102 First Street South Kalona IA 52247 USA WWW CIVCO COM Sales info civco com Customer Service info civco com Technical Support service civco com Phone 800 445 6741 ...

Page 2: ...eavier than 50 pounds use two or more people to lift the load Do not use this system until it has been properly installed Use the device only for the intended use as described in this manual Explosion hazard do not use in the vicinity of flammable gases or liquids Do not operate the device if it is not working properly or if any part of the device has been dropped or damaged For use only with CIVC...

Page 3: ...m is required Incoming water pressure 30 50 psi 207 345 kPa Recommended minimum flow rate 1 gpm 3 8 lpm at 30 psi 207 kPa Temperature Cold water supply 50 80 F 10 27 C Drain Requirements A drain connection is required using supplied tubing the device may drain directly into an existing sink dishwasher drain tailpiece or connected permanently to existing plumbing drain Environmental Requirements In...

Page 4: ...embly Small 610 2000 1 HLD Bottle Cap Assembly Large 610 2306 1 Power Cord A7061 01 1 Probe Connector Holder A1064 01 2 Probe Connector Holder Pad back A9064 01 2 Probe Connector Holder Pad base A9065 01 2 EZ Anchor Kit A6164 01 6 Wall Bracket A1053 01 1 Water Supply and Drain Tubing Assembly A0304 01 1 Operators Manual M0066 1 Installation Manual this document M0065 1 ASTRA VR Quick Start Guide 5...

Page 5: ...et Failure to follow these installation instructions may affect the operation and effectiveness of this device Figure 1 below shows a typical installation Figure 1 Typical ASTRA VR Installation The on off switch power cord vent water supply and drain are located on the right side facing of the device These connections may be routed behind the unit if your plumbing or electrical outlet is located t...

Page 6: ...r inlet hoses and connectors and a 10 ft removable hospital grade power cord Use of the supplied power cord is required to maintain proper grounding and safe operation of the device This power cord is required to be plugged directly into a GFCI receptacle Figure 2 Mounting hole locations and overall space requirements indicated by dotted lines ...

Page 7: ...t the load C Mounting Probe Connector Holders Two probe connector holders A1064 01 are supplied with the device and should be installed on the left side of the unit refer to Figure 2 for mounting locations Install probe connector holders with the EZ Anchor kit A6164 01 supplied Once mounted to the wall attach probe connector holder pads back A9064 01 and base A9065 01 in each holder D Plumbing Ins...

Page 8: ...ste drain This can be into a typical dishwasher type tailpiece under a sink a floor standpipe or directly into a waste sink secure line to prevent drain water from splashing Ensure any required air gaps are maintained Figure 4 Drain and water quick disconnect fittings E Electrical Connection It is the installer s responsibility to ensure the device is installed in a manner which is compliant with ...

Page 9: ...ting and stop valve within 8 ft o Water supply pressure within 30 50 psi o Water supply flow rate greater than 1 gpm o Water supply temperature within 50 80 F Drain available within 8 ft Physical Mounting bracket and probe holders installed Minimum clearance requirements met Environmental Environmental requirements met Site Photograph Photograph attached of intended installation site Other All par...