ASTRA® VR Endovaginal/Endorectal Probe Reprocessor
Installation Manual
M0065_C
2
ASTRA® VR
3.
SITE REQUIREMENTS & SPECIFICATIONS
Actual Size (ASTRA VR):
Width:
12.0” (30.5 cm)
Depth:
14.0” (35.6 cm)
Height:
53.0” (134.6 cm)
Weight:
73 lbs. (33.1 kg)
Minimum Clearances Required:
Front:
12.0” (30.5 cm)
Left:
12.0” (30.5 cm)
Right:
12.0” (30.5 cm)
Top:
6.0” (15.2 cm)
Electrical Requirements:
•
Service:
115 VAC (±10%) 60Hz, 15A Hospital Grade GFCI receptacle within 8 ft. (2.4 m)
•
Fuse:
2.0 Amp Slo Blo, 250V, 5mm X 20 mm
•
Category II installation
Water Supply Requirements:
•
A permanent cold-water supply connection within 8 ft. (2.4 m) is required.
•
Incoming water pressure: 30 – 50 psi (207 – 345 kPa)
•
Recommended minimum flow rate: 1 gpm (3.8 lpm) at 30 psi (207 kPa)
•
Temperature: Cold water supply 50 - 80 °F (10 – 27 °C)
Drain Requirements:
•
A drain connection is required using supplied tubing; the device may drain directly into an
existing sink,
dishwasher drain tailpiece
, or connected permanently to existing plumbing
drain.
Environmental Requirements:
•
Indoor use only
•
Altitude up to 6560 ft. (2000m)
•
Operating Temperature: 65 - 80°F (18 - 27°C)
•
Pollution degree 2 environment
•
Room Humidity 20 - 80% RH non-condensing
•
When disposing of this equipment and accessories, follow all applicable national, state and local
regulations and guidelines.
•
Ventilation should meet local / hospital standards for any area handling liquid high-level
disinfectants.