ASTRA® VR Endovaginal/Endorectal Probe Reprocessor  

Installation Manual 

M0065_C

 

ASTRA® VR 

 

3.

 

SITE REQUIREMENTS & SPECIFICATIONS 

 

 Actual Size (ASTRA VR): 

Width: 

12.0” (30.5 cm)

 

 

 

Depth: 

14.0” (35.6 cm)

 

 

 

Height: 

53.0” (134.6 cm)

   

 

Weight: 

73 lbs. (33.1 kg)

 

 

 

 

Minimum Clearances Required: 

Front:

 

12.0” (30.5 cm)

   

 

Left:

 

12.0” (30.5 cm)

   

 

Right:

 

12.0” (30.5 cm)

   

 

Top:

 

6.0” (15.2 cm)

   

 

 

Electrical Requirements:  

• 

Service:  

115 VAC (±10%) 60Hz, 15A Hospital Grade GFCI receptacle within 8 ft. (2.4 m) 

• 

Fuse:  

2.0 Amp Slo Blo, 250V, 5mm X 20 mm   

• 

Category II installation

 

 

Water Supply Requirements: 

• 

A permanent cold-water supply connection within 8 ft. (2.4 m) is required. 

• 

Incoming water pressure: 30 – 50 psi (207 – 345 kPa)  

• 

Recommended minimum flow rate: 1 gpm (3.8 lpm) at 30 psi (207 kPa) 

• 

Temperature: Cold water supply 50 - 80 °F (10 – 27 °C)

 

 

Drain Requirements:  

• 

A drain connection is required using supplied tubing; the device may drain directly into an 
existing sink, 

dishwasher drain tailpiece

, or connected permanently to existing plumbing 

drain. 

 

Environmental Requirements:  

• 

Indoor use only 

• 

Altitude up to 6560 ft. (2000m)   

• 

Operating Temperature: 65 - 80°F (18 - 27°C) 

• 

Pollution degree 2 environment 

• 

Room Humidity 20 - 80% RH non-condensing 

• 

When disposing of this equipment and accessories, follow all applicable national, state and local 

regulations and guidelines. 

• 

Ventilation should meet local / hospital standards for any area handling liquid high-level 

disinfectants. 

 

 

 

Summary of Contents for Astra VR

Page 1: ... registered trademark of CIVCO Medical Solutions All other trademarks are property of their respective owners Printed in the USA CIVCO Medical Solutions 102 First Street South Kalona IA 52247 USA WWW CIVCO COM Sales info civco com Customer Service info civco com Technical Support service civco com Phone 800 445 6741 ...

Page 2: ...eavier than 50 pounds use two or more people to lift the load Do not use this system until it has been properly installed Use the device only for the intended use as described in this manual Explosion hazard do not use in the vicinity of flammable gases or liquids Do not operate the device if it is not working properly or if any part of the device has been dropped or damaged For use only with CIVC...

Page 3: ...m is required Incoming water pressure 30 50 psi 207 345 kPa Recommended minimum flow rate 1 gpm 3 8 lpm at 30 psi 207 kPa Temperature Cold water supply 50 80 F 10 27 C Drain Requirements A drain connection is required using supplied tubing the device may drain directly into an existing sink dishwasher drain tailpiece or connected permanently to existing plumbing drain Environmental Requirements In...

Page 4: ...embly Small 610 2000 1 HLD Bottle Cap Assembly Large 610 2306 1 Power Cord A7061 01 1 Probe Connector Holder A1064 01 2 Probe Connector Holder Pad back A9064 01 2 Probe Connector Holder Pad base A9065 01 2 EZ Anchor Kit A6164 01 6 Wall Bracket A1053 01 1 Water Supply and Drain Tubing Assembly A0304 01 1 Operators Manual M0066 1 Installation Manual this document M0065 1 ASTRA VR Quick Start Guide 5...

Page 5: ...et Failure to follow these installation instructions may affect the operation and effectiveness of this device Figure 1 below shows a typical installation Figure 1 Typical ASTRA VR Installation The on off switch power cord vent water supply and drain are located on the right side facing of the device These connections may be routed behind the unit if your plumbing or electrical outlet is located t...

Page 6: ...r inlet hoses and connectors and a 10 ft removable hospital grade power cord Use of the supplied power cord is required to maintain proper grounding and safe operation of the device This power cord is required to be plugged directly into a GFCI receptacle Figure 2 Mounting hole locations and overall space requirements indicated by dotted lines ...

Page 7: ...t the load C Mounting Probe Connector Holders Two probe connector holders A1064 01 are supplied with the device and should be installed on the left side of the unit refer to Figure 2 for mounting locations Install probe connector holders with the EZ Anchor kit A6164 01 supplied Once mounted to the wall attach probe connector holder pads back A9064 01 and base A9065 01 in each holder D Plumbing Ins...

Page 8: ...ste drain This can be into a typical dishwasher type tailpiece under a sink a floor standpipe or directly into a waste sink secure line to prevent drain water from splashing Ensure any required air gaps are maintained Figure 4 Drain and water quick disconnect fittings E Electrical Connection It is the installer s responsibility to ensure the device is installed in a manner which is compliant with ...

Page 9: ...ting and stop valve within 8 ft o Water supply pressure within 30 50 psi o Water supply flow rate greater than 1 gpm o Water supply temperature within 50 80 F Drain available within 8 ft Physical Mounting bracket and probe holders installed Minimum clearance requirements met Environmental Environmental requirements met Site Photograph Photograph attached of intended installation site Other All par...

Reviews: