ASTRA® VR Endovaginal/Endorectal Probe Reprocessor
Installation Manual
M0065_C
1
ASTRA® VR
1.
INSTALLATION MANUAL
This instruction manual contains important information to install this device safely and
effectively. Review this manual before installation. Keep this manual in a safe and accessible
location. Before beginning this installation, make certain that you have read and understood
these instructions.
It is the installer’s responsibility to ensure the equipment is installed in a manner which
conforms to any state/local building code requirements. This installation requires that
appropriate plumbing and electrical supplies are present before connecting the ASTRA VR
system. Installing the ASTRA VR system does not require specialized training. Contact CIVCO
Medical Solutions with any questions or comments regarding the information contained in this
manual at (800) 445-6741 or via email at [email protected].
2.
WARNINGS & CAUTIONS
CAUTION:
•
Read the entire manual before using the device.
•
No user serviceable parts inside. Shock hazard may be present behind access panel.
Panel removal should be performed only by qualified service personnel.
•
The device weight exceeds 50lbs (22.7kg). When lifting loads heavier than 50
pounds, use two or more people to lift the load.
•
Do not use this system until it has been properly installed.
•
Use the device only for the intended use as described in this manual.
•
Explosion hazard – do not use in the vicinity of flammable gases or liquids.
•
Do not operate the device if it is not working properly or if any part of the device
has been dropped or damaged.
•
For use only with CIVCO Medical Solutions approved high-level disinfectants.
•
Connect the system to a hospital grade GFCI receptacle only. GFCI outlet must be
tested on a regular basis per the GFCI manufacturer’s recommendation.
•
Do not operate system if fluid leakage is present, contact CIVCO Medical Solutions.
•
The device is pre-programmed for use with a specific High-Level Disinfectant (HLD),
determined by the user.
•
Failure to properly lock the quick connect connectors into place may result in leaks.
•
Use of power cords, drain and water inlet hoses and connectors not supplied by
CIVCO Medical Solutions may affect the safety and performance of this device.
•
Do not use abrasive cleaning materials, solvents, or alcohol to clean the device.
•
Do not pour any items directly into disinfection chamber.
•
Do not use this device for any purpose other than its intended use.
CAUTION: Hot Surface
•
When cabinet door is opened, use caution when removing and installing HLD
bottles. The HLD warmer plate surface temperature may reach 161°F (72°C).