The optimum location
The correct location makes an important contribution to the accuracy of the weighing
results of high-resolution analytical and precision balances.
Hence, ensure a stable, vibration-free position as horizontal as possible.
Avoid
• Direct sunlight
• Excessive temperature fluctuations,
• Drafts (Power ----- Air Conditioning System, Fans can also cause drafts)
The best position is an a stable bench in a corner protected against drafts as far
possible from doors, windows, radiators or the ventilation slots of air conditioners.
Anti-theft device
Citizen Balance are equipped with a lug for optional anti-theft device.
The anti-theft device (cable with lock) is suitable for all models. It is available from
CITIZEN under order number CAD01.
2.1 Location
12
11
•
Place the components listed below inside the chamber in the order given:
(For CG Series Balance)
– Bush
– Weighing pan
•
Balances with a Rectangular Weighing Pan and a Weighing Capacity over 10 kg
– Place the weighing pan on the balance/scale
Under Weighing
A port for a under weighing hanger is located on the bottom of the balance.
• Remove the under weighing cover from the bottom of the balance (as shown
alongside)
• Suspend the below-balance weighing hook supplied from the hanger
• If necessary, install a shield for protection against drafts