85
Note:
ð
It is assumed that the user has already performed the above procedure and the s c a l e
is in F PiP mode.
ð
When the pipette calibration procedure process has not started the system will
perform normal weighing with the unit as gm.
Calibration procedure:
ð
The calibration procedure repeats itself 'n' times where 'n' is the count entered in the
count cycle settings (in FPiP settings).
ð
Once in F PiP mode, press the <CAL/MENU> key to start the calibration process.
ð
Once the <MENU/CAL> key is pressed, it will ask for the initial volume
V
0.
ð
For a calibration procedure of 'n' times, the scale will ask for
V
'n' times. Each of t h e s e
0
value will be stored as 'PLACE n'.
st
ð
Thus, for the 1 time, place the initial volume and press <CAL/MENU> key once the
stability is achieved. This is accepted as 'PLACE 1'.
ð
Follow the above procedure for 'PLACE 1' to 'PLACE n'.
ð
Repeat the above procedure for half volume V and full volume V each 'n' times
half
full
respectively.
ð
In case of a successful calibration the scale will display 'PiP done'. If not, then it will
ask to repeat the step wrongly performed.
Volume Settings:
Initial volume V
0:
ð
Press <TOGGLE> key to enter F PiP settings.
ð
Select 'V ' in F PiP settings by pressing <TOGGLE> key.
0
ð
Press < TARE> to enter the 'V ' settings.
0
ð
The user can change the volume with the help of <CAL/MENU> key and <TOGGLE>key.
ð
Pressing the <CAL/MENU> key will shift the cursor to the right in a cyclic way and pressing
the <TOGGLE> key will increment the digit.
ð
Press the <TARE> key to select the value. This will bring you back to the F PiP menu
ð
The default volume is,
V = 10.00 uL
0
Half volume V
Half :
ð
Press <TOGGLE> key to enter F PiP settings.
ð
Select '
V
' in F PiP settings by pressing <TOGGLE> key.
Half
ð
Press < TARE> to enter the '
V
' settings.
Half
ð
The user can change the volume with the help of <CAL/MENU> key and <TOGGLE>key.
ð
Pressing the <CAL/MENU> key will shift the cursor right in a cyclic way and pressing the
<TOGGLE> key will increment the digit.
ð
Press the <TARE> key to select the value. This will bring you back to the F PiP menu
ð
The default volume is,
V = 11.00 uL
Half
Full volume, V
Full. :
ð
Press <TOGGLE> key to enter F PiP settings.
ð
Select 'V ' in F PiP settings by pressing <TOGGLE> key.
Full.
ð
Press < TARE> to enter the 'V ' settings.
Full.
ð
The user can change the volume with the help of <CAL/MENU> key and <TOGGLE>key.
ð
Pressing the <CAL/MENU> key will shift the cursor right in a cyclic way and pressing the
<TOGGLE> key will increment the digit.
ð
Press the <TARE> key to select the value. This will bring you back to the F PiP menu
ð
The default volume is,
V = 12.00 UL.
Full
86