V1.0I2 PULSE OXIMETER OPERATOR
'
S MANUAL
24
d
)
During operation please note:
After operation the oximeter, you should take off the probe and make daily
maintenance. If you will not use it for a long time please take out battery.
e
)
Battery maintenance
Please take out battery if you will not use the oximeter for a long time.
Please charge the battery fully if you will not use it for a long time.
Please charge over 14 hours at first time, or will reduce the battery life.
If occurs any abnormal phenomena, you should stop using immediately and
reuse after inspection by a technologist.
7.2 Calibration and Verification
The performance should be checked every year. It should also be checked
after every maintenance and repair.
Required Test Equipment: SpO
2
signal Simulator
Note:
The simulator cannot be used to assess the accuracy of a pulse
oximeter probe or a pulse oximeter.
7.2.1 Control Key Verification.
Press Menu key, display the history data.
7.2.2 Sound Verification
a)
.
Set the oximeter sound ON.
b)
.
The simulated heart beep sound Can be heard.
7.2.3 SpO
2
& Pulse Rate Measurement Value Verification
a)
.
Connect SpO
2
Probe to the SpO
2
connector on the oximeter.
b)
.
Insert the operator’s finger into the finger sensor, the SpO
2
measured value
of a healthy person should be from 95% to 99%, and the pulse rate is
same as heart rate,
c)
.
If SpO
2
Simulator is available, verify the accuracy of Oxygen Saturation
Value with BCI probes as follows:
Oxygen Saturation Tolerance
96% ±2%
86% ±2%
70% ±3%