V1.0I2 PULSE OXIMETER OPERATOR
'
S MANUAL
17
CHAPTER 6 Other Settings
6.1 Alarm Setting
Press the
ENTER
button to enter into the main menu and the
“1. Alarm Set”
item is selected automatically, and then press the
ENTER
button to enter into
the Alarm Set submenu.
1) SpO
2
alarm setup
After you enter into the
“Alarm Set”, the highlimit value below SpO
2
item is
selected automatically (refer to Fig.17), and then press the
ENTER
button to
highlight the value (refer to Fig.18). Press the
UP
button or the
DOWN
button
to adjust the value, and then press the
ENTER
button to finish the setup of
SpO
2
% upper alarm limit. The range of upper limit is 70-99.
Fig.17 Fig.18
In the Alarm Set submenu, press the
UP
button
or the
DOWN
button to select
the lowlimit value below SpO
2
item(refer to Fig.19), and then press the
ENTER
button to highlight the value(refer to Fig.20). Press the
UP
button or the
DOWN
button to adjust the value, and then press the
ENTER
button to finish
the setup of SpO
2
% lower alarm limit. The range of lower limit is 69-98.