![Chemtrac HydroAct HA4 User Manual Download Page 38](http://html1.mh-extra.com/html/chemtrac/hydroact-ha4/hydroact-ha4_user-manual_2595954038.webp)
38
Chapter 3. Sensors
3.1.1.5
Technical Data
UV254 sensor profile view
1. 430 mm
2. 365 mm
3. 195 mm
Range
0..5 UVA, 0..120% UVT
Accuracy
±
0.5% FS
Resolution
0.001 UVA.
Path Length
2mm, 4mm, or 10mm (to be specified at the time of order)
Sampling Time
10 seconds.
Flow Rate
300..1000ml/min
Pressure rating
20 PSI
Cleaning
Automatic cleaning option (to be specified at the time of order)
Humidity Control
Humidity sensor with large regenerable desiccant system
Light Source
Low pressure mercury UV lamp
Lamp life
2 year expected
Enclosure
IP 65 (NEMA 4X), wall mountable
Fluid Connections
1/4" push-fit inlet/outlet.
Storage Temperature
-20 to 60°C
Operating Temperature
0 to 45°C
Rated Water Temperature
75°C
Weight
10kg