Doc.No: 994-2025 v. 1.0 · Issue date: 13-Dec-2019
Template 880-0011-79B v.3
Page 1 of 4
ChemoMetec A/S · Gydevang 43 · 3450 Allerod · Denmark · [email protected] · www.chemometec.com ·
Technical Note No. 2025 Rev. 1.0
NucleoCounter® NC-202™
NucleoCounter® NC-202™ Performance Qualification
Instrument and Materials Specifications
Purpose:
To specify the instrument, test kit and cassettes being used for the performance qualification, together
with the software and firmware installed at the time of performing the qualification.
Instrument Type:
NucleoCounter® NC-202™ (Cat. no. 900-2020)
Instrument Serial Number
Firmware versions
System Controller
Piston
Software version
NC-202
PQ Kit
(Cat. no. 912-2022)
Expiry Date and Lot number
Via2-Cassette™ (Cat. no. 941-0024)
Expiry Date and Lot number
Zero Count Procedure
□
N/A
Purpose
:
To ensure that the entire NC-202™ system counting performance is successful when no cassette is
present.
Select the
IQ/OQ
protocol in the drop-down menu, or via the
Select Protocol
icon
□
OK/Yes
□
Fail/No
Execute the protocol by pressing the RUN button on the front of the instrument and
select
Zero Count
in the message box.
□
OK/Yes
□
Fail/No
Zero Count successful as indicated by Passed.
□
OK/Yes
□
Fail/No
Page 53 of 127
Summary of Contents for NucleoCounter NC-202
Page 1: ...997 0005 NucleoCounter NC 202 Document Compilation V 1 0 Page 1 of 127...
Page 14: ...Page 14 of 127...
Page 15: ...Page 15 of 127...
Page 16: ...Page 16 of 127...