Doc.No: 992-2023 v. 1.0 · Issue date: 05-Dec-2019
Template 880-0011-79C v.1
Page 1 of 1
ChemoMetec A/S · Gydevang 43 · 3450 Allerod · Denmark · [email protected] · www.chemometec.com ·
912-2022
NC-202
PQ Kit
Contents
P/N: 071-0029
P/N: 071-0030
P/N: 071-0031
One vial with 1.5 mL of fluorescent 10 µm beads, low bead concentration.
One vial with 1.5 mL of fluorescent 10 µm beads, medium bead concentration.
One vial with 1.5 mL of fluorescent 10 µm beads, high bead concentration.
Buffer solution: 0.02% Tween-20 and 2mM ~ 0.01% NaN
3
in H
2
O.
CAS no. for Tween-20 is 9005-64-5
CAS no. for NaN
3
is 26628-22-8
EINECS no. for NaN
3
is 247-852-1
Application
NC-202
PQ Kit
is used for performing a Performance Qualification (PQ) of the NucleoCounter®
NC-202™
.
The test kit is not a counting standard.
Principle
NC-202
PQ Kit
contains multicoloured beads that allow testing of the performance of the counting
capabilities of the NucleoCounter® NC-202
™.
The vials contain various concentrations of two
types of beads. All vials contain beads that will be detected in the AO and DAPI channel. Both
bead types will contribute to the total counts.
Usage
With respect to the description of the detailed PQ procedure, please refer to the appropriate
technical notes, certificate and user manuals for the NucleoCounter® NC-202
™
instruments and
the NC-View™ Software.
NC-202
PQ Kit
is for research and development purposes only and is not for diagnostic or
therapeutic use.
Storage
The fluorescent bead solution should be stored at 2-8°C. Protect against light.
Stability
The shelf life for the kit is 15 months from the production date. The expiry date is shown on the kit
as well as on the bead vial label.
Safety
Information
Fluorescent bead solution: In case of contact with eyes, rinse immediately with plenty of water
and seek medical advice.
If a spill is observed perform a clean-up of the area, which may have been in contact with the
solution.
Use gloves and suitable protective clothing.
Please also refer to SDS regarding safety information.
Disposal of
Waste
After use, the
NC-202
PQ Kit
should be disposed of according to national or local laws and
regulations regarding the nature of the mixture it contains.
Page 39 of 127