2 Introducing the NucleoCounter
11
A fluorescent image of the content of the measurement chamber is thus . Example
of such an image is shown in Figure 4. Each spot in the image represents a cell
containing DNA stained with PI.
Figure 4
. Fluorescent image (from NucleoView PC software) of cell DNA contained
in the measurement chamber of a NucleoCassette
2.1.2 Built-in image analysis
The NucleoCounter is equipped with advanced software for image analysis. This
integral software analyzes the recorded image and counts the num ber of nuclei in
the image. As the volume of the cassette measurement chamber and the dilution of
the lyzate is known, the cell density in the initial sample can be calculated. The
concentration of cells in the initial sample is presented in the NucleoCoun ter
display. If the optional software (NucleoView) is used, the results and the recorded
fluorescent image can be viewed.
The entire process takes approximately 30 seconds, from pressing “ Run” to the
presentation of the result in the display of the NucleoC ounter.
Performing a cell count with the NucleoCounter is objective and the results are
independent of the operator.
2.2
The NucleoCassette
The NucleoCassette is a custom made disposable plastic device designed for
optimal sample handling and safe disposal. The cassette is shown in figure 5. The
main features of the NucleoCassette are the piston, the flow system containing the
fluorescent dye, propidium iodide, and the measurement chamber.
Summary of Contents for 900-0300
Page 1: ...NucleoCounter YC 100 User s Guide Revision 1 3 Technology that counts...
Page 2: ...i This page was intentionally left blank...
Page 4: ...iii This page was intentionally left blank...
Page 6: ...v This page was intentionally left blank...
Page 8: ...vii This page was intentionally left blank...
Page 10: ...ix This page was intentionally left blank...
Page 12: ...Introduction and intended use xi This page was intentionally left blank...
Page 20: ...1 Inspection and Unpacking of Equipment 8 This page was intentionally left blank...
Page 26: ...Fejl Henvisningskilde ikke fundet 14 This page was intentionally left blank...
Page 44: ...7 Maintenance of NucleoCounter 32 This page was intentionally left blank...
Page 48: ...8 Troubleshooting Error messages 36 This page was intentionally left blank...
Page 55: ...43 2 1 3 Metal E nclosure Figure 36 Printer Output connector seen from the cable entry side...