8 Troubleshooting - Error messages
33
8 Troubleshooting - Error messages
Under certain conditions the NucleoCounter will display error messages du ring
operation. Corrective actions are suggested below but if they do not correct the
errors, contact ChemoMetec A/S or the local distributor.
8.1
No valid cassette
When trying to analyze a cassette, which is not valid, the NucleoCounter displays
the error message shown in Figure 28.
No valid
cassette
Press any key
Figure 28.
The error message displayed when the NucleoCounter does not
recognize the cassette as valid
This error message is displayed in the following situations:
The cassette has been analyzed previously
The cassette is not inserted properly into the NucleoCounter
There is no cassette inserted into the NucleoCounter
8.2
Analysis aborted
The NucleoCounter displays the error message show n in Figure 29 when the
analysis has been aborted because the stained mixture inside the cassette has not
reached the chamber within a given time.
Analysis aborted
Flow sensor error
Refer to manual
Press any key
Figure 29.
Error message displayed by the NucleoCounter when the analysis has
been aborted
A common cause of this is that an insufficient volume of lyzate mixture has been
loaded into the cassette. Load a new cassette and repeat the ana lysis.
The error will also occur if the cassette has not been inserted properly into the
NucleoCounter. Then the instrument cannot press the piston down and the error
occurs. Take out the cassette, insert it properly and run the analysis.
Summary of Contents for 900-0300
Page 1: ...NucleoCounter YC 100 User s Guide Revision 1 3 Technology that counts...
Page 2: ...i This page was intentionally left blank...
Page 4: ...iii This page was intentionally left blank...
Page 6: ...v This page was intentionally left blank...
Page 8: ...vii This page was intentionally left blank...
Page 10: ...ix This page was intentionally left blank...
Page 12: ...Introduction and intended use xi This page was intentionally left blank...
Page 20: ...1 Inspection and Unpacking of Equipment 8 This page was intentionally left blank...
Page 26: ...Fejl Henvisningskilde ikke fundet 14 This page was intentionally left blank...
Page 44: ...7 Maintenance of NucleoCounter 32 This page was intentionally left blank...
Page 48: ...8 Troubleshooting Error messages 36 This page was intentionally left blank...
Page 55: ...43 2 1 3 Metal E nclosure Figure 36 Printer Output connector seen from the cable entry side...