97
mode
2 kV common
mode
differential mode
2 kV common
mode
network that supplies buildings used for
domestic purposes, commercial or
hospital, clinic environment.
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
.
<5 %UT
(>95 %dip in UT)
for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5 cycles
<5 %UT
70 %UT
(30 %dip in UT)
for 25 cycles
<5 %UT
<5 %UT
(>95 %dip in UT)
for 5 s
<5 %UT
(>95 %dip in
UT) for 0,5 cycle
40 %UT
(60 %dip in UT)
for 5 cycles
<5 %UT
70 %UT
(30 %dip in UT)
for 25 cycles
<5 %UT
<5 %UT
(>95 %dip in
UT) for 5 s
Mains power quality should be that of a
typical public low-voltage power supply
network that supplies buildings used for
domestic purposes, commercial or
hospital, clinical environment.
If the user of the STARLING SV requires
continued operation during power mains
interruptions; it is recommended that
STARLING SV be powered from a
separate power supply (UPS, etc.).
Power frequency
(50/60 Hz)
magnetic field
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
public low-voltage power supply network
that supplies buildings used for domestic
purposes, commercial or hospital, clinic
environment.