7.2
Warning Messages
The following warning messages may be encountered in the course of normal system
operation. The following instructions are intended to assist the user in dealing with
them.
1. Diagnosis:
CO is extremely high and a pacemaker is active on the patient
Cause:
The patient has a Pacemaker which can result in interruption in the Starling SV
signal.
Solution:
Usually Pacemaker activity will not affect the Starling SV signal unless it is a
Unipolar Pacemaker or External Pacemaker activities which can result in distorted
Starling SV signal.
If there is a relatively high CI reported by the system, verify that the Pacemaker
lead distance from the Starling SV sensor is no less than 2.5 inches. If not, replace
the Starling SV Sensors with new sensors and make sure the sensor is positioned
more than 2.5 inches from the lead.
2. Message:
Testing has not been completed on the MRI compatibility of either the sensor
or the monitor.
Reason – The battery supply is low and is under 10% of its capacity.
Cause:
The device was not charged for a long period of time.
Solution:
Connect the device to Main power supply as soon as possible.
7.3
Other Problems
1. Problem:
One or more channels do not display correctly.
Cause:
1. Sensors are not properly connected.
2. Patient movement during the measurement process.