Specifications
Skin Voltage Tolerance:
±300 mV
Frequency Response:
8Hz
Time Constant:
0.125 Sec
CMRR:
70db
Filter:
75KHz
Safety Requirements:
According to IEC/EN 60601-1 standard. Class I
Equipment BF type applied part.
Rated Voltage/Current:
Power switch supply 100V to 240V
Rated frequency
50/60 Hz.
Fuses:
2 X T2AL250V
A/D
converter at 12 bit (2.5 mV LSB).
Sample rate
500 per sec.
Features
Hemodynamic display options
Height/Weight units
Dedicated Starling SV Electronic Medical Records
interface (XlS PDF & XML file formats, Password
Protection)
Hemodynamic status report with visit-to-visit data
Clinical Protocol Wizards
Event Management
Touch Screen control
Serial and Ethernet connections
Ambient Temperature and Humidity:
Operation condition:
10-40°C/ 50-104°F, 50-75% RH
Atmospheric pressure:
700hPa to 1060hPa
Storage condition:
0-50°C/ 32-122°F, 50-75% RH
Delivery condition:
0-50°C/ 32-122°F, 50-75% RH
Technical Characteristics of Acquisition Module
Isolated preamplifier in accordance with IEC 60601-1 standard.
Note
The Starling SV is not defibrillator proof.