Figure 4.1
– General Options screen with Settings pop-up menu
4.1
Height Units
This option allows user to set the device to work according to the Metric or USA
system. Moving from one system to another changes all measurable units from cm
units to inch units, and vice versa.
4.2
Weight Units
This option allows user to set the device to work according to the Metric or USA
system. Moving from one system to another changes all measurable units from kg
units to lb and vice versa.
4.3
Electric Grid
The STARLING SV filters away interferences caused by the electric grid. The electric
grid emits signals with frequency of 60 Hz in the US and 50 Hz in Europe and most
of the rest of the world. To customize the filter to your geographic location select 60
Hz if you are located in the USA and 50 Hz if located elsewhere. If you are not sure
about the local grid system parameters please contact your IT personnel or electricity
provider.
CAUTION
Choosing an incorrect grid system might result in signal interference
problems which can lead to suboptimal performance.
4.4
Audible Alarm
Change this field to Off to avoid low battery audible alerts. The buzzer status icon will
appear on the top left side of the run bar of the display screen (Run, Dashboard,
Trend and Numeric display).
4.5
Date Format
This option allows the user to set the date display format in the device to show either
Day/Month/Year or Month/Day/Year.
4.6
Time Format
This option allows the user to set the time format in the device to 24 hrs or 12 hrs