Operator manual Cube 13-00170-US Rev C
Page 43 of 51
Wireless foot control is not working.
Check the battery of the wireless footswitch.
Check the registry of the wireless footswitch.
Check if the wireless footswitch is chosen in the set-up submenu.
If you cannot solve the malfunctioning, switch off the laser and contact an Authorized Service
Center.
8.3 Safety check
The following safety checks must be performed every 24 months by a qualified service engineer.
Visual inspection of the unit and its accessories for mechanical damage that might impair
operation;
General function check;
Check of the visual and audible indicators;
Check the electrical power;
Check the power emission of the laser.
8.4 Technical support, repair and testing
In case technical support, contact the following address:
UNITED STATES:
T: 1-800-494-3395 USA
Email:
Before sending the device to the Service Center, it is required to disinfect the device and
accessories as written in the present manual.
The device may be sent in for repair or for safety inspection only in its original packaging, including
all accessories.
This is the only provider of technical service for its products, unless otherwise indicated by the
C
ompany. This is to guarantee the product’s safety and for the preservation of the product’s
characteristics.