Operator manual Cube 13-00170-US Rev C
Page 27 of 51
Notice
Treatment time and Total Joule are automatically calculated by the device, and cannot be set
by the user.
To manage the parameters select the field that will be highlighted with a green square and then
press:
Minus / Plus key or
to enter in a Keyboard to implement the modification.
Treatment total time. This value is automatically calculated by
the device.
The device automatically calculates the energy used during
treatment (in J) from the power values and the selected laser
activation time. Total energy during application is displayed.
It is adjustable between 0.1 W and max value. In 0.01 W
increments by typing in the numbers or by touching
.
The following power:
Cube 3: CW power 0.1
– 12 W;
Cube 4: CW power 0.1
– 15 W;
Notice
In ISP modality power emission is automatically set.
The device calculates the average power from the power
values and duty cycle.
Notice
In ISP modality average power can be set in base of
model.
Frequency is adjustable between CW, 1 - 20000 Hz or ISP.
It is adjustable by typing in the numbers or touching
Increments are adjustable as follows:
from 1 to 20000 Hz
– step of 1 Hz.
Time of the each phase is adjustable between 1 to 999 s.
It is adjustable by typing in the numbers or by touching
Wavelengths are adjustable based on the model, they can be
selected in single or combined modality.
Set the parameters for a phase and proceed for the maximum number of 12 phases.
To delete completely a phase, select it and then press delete phase.
To confirm the personal treatment press the save button and enter the name.
After name's confirmation, the device will store the treatment.
Total Joule
Power
Frequency
Wavelength
Phase time
Average power
Treatment time