CGEO-PR-VW Vibrating Wire Portable Readout
CGEO INTERNATIONAL LIMITED
Installation Manual
another gauge. Repeat the steps, if the result is ok, it shows that the first gauge is not
operational; and if still doesn't work consult the factory to schedule return and repair of
Readout Box.
4.04.4 The Reading is unstable or no reading
Improper DISPLAY MODE switch may be selected. Please change, and
check in the order of D
→
E
→
F
→
A
→
B
→
C, and if there is stable reading in some
MODE, just stop changing.
4.04.5 RS232 is not operational
Please check the following items
:
a.) The communication format of the software is 8 data (no odd or even), 1 stop bit,
9600 bps;
b.)The hardware of RS232, including connection lead Export/Import;
c.) The length of connecting lead is no longer than 10m. If the above checks do not
resolve the situation, please consult the factory to schedule return and repair.