CETONI Nemesys S Hardware Manual
21
5.3
Dead volume free syringe holder
The dead volume free syringe holder is included in the scope of delivery of the CETONI Nemesys S and
can be mounted instead of the Contiflow Valve (see 5.10). It is particularly suitable for use with small
syringes when a continuous fluid flow is not required. In this case, the syringe is screwed directly against
the downstream capillary. You can use both, flanged capillaries and fittings with ferrule. The following
sectional view illustrates the principle:
The syringe is mounted in the same way as on the Contiflow valve, which is described below:
5.4
Mounting a Syringe
The following procedure describes the mounting of the SETonic glass syringes specially optimized for the
Nemesys S (for more information, see 5.13). If you would like to use other syringes, please refer to the
information on the Universal Syringe Holder Nemesys S under 5.12.
Before you mount a syringe on the CETONI Nemesys S, it must be selected in the software. The SETonic
glass syringes are already preconfigured. The procedure is described in the software manual.
Proceed as follows to mount a syringe on the CETONI Nemesys S
(1)
Push the syringe piston completely into the syringe.
(2)
Use the software to bring the piston holder into a position that allows the syringe to be screwed into
the syringe holder.