34
CETONI Nemesys S Hardware Manual
5.13
Glass Syringes - Instructions for Use
Only special SETonic glass syringes may be used with the CETONI Nemesys S. These syringes are matched
to the CETONI Nemesys S with regard to their dimensions. The use of other syringes can lead to collisions
and thus to damage to the syringe or the device.
You recognize these syringes by a CET label on the front cap.
Only with the Universal Syringe Holder Nemesys S (see section 5.12) the use of other syringes is
permitted, whereby the reliable and safe function in this case must be ensured by the operator and cannot
be guaranteed by CETONI GmbH.
ATTENTION
. Unless you use the Universal Syringe Holder Nemesys S, only the
SETonic glass syringes marked CET may be used with the CETONI Nemesys S. The use
of other syringes can lead to collisions and thus to damage to the syringe or the
device.
CETONI is not the manufacturer of the syringes. However, with the following information we would like
to support you as much as possible in handling the syringes.
•
Check the syringes for cracks before using.
•
To remove air bubbles, aspirate the sample and dispense it quickly, while leaving the tip of the
syringe in the liquid. You can also hold the syringe up, so that the air bubble rises and then
dispense some of the sample. Operate the syringes with the outlet pointing upwards to improve
venting. (See 5.6)