CETONI Nemesys S Hardware Manual
33
(5)
Place the syringe on the lower part of the syringe holder
1
and push the syringe piston against the
base body
2
of the piston holder. Then rotate the syringe down holder
3
over the syringe, slide it
down until it rests on the syringe and tighten the clamping screw
4
until the syringe is securely
clamped.
(6)
Insert the adapter plate
5
(if required) over the syringe piston and tighten the clamping screw
6
of
the piston holder until the piston is fixed.
(7)
Finally, push the counter holder
7
against the syringe and fix the position with the clamping screw
8
.
The counter holder prevents the syringe from slipping at higher pressures.
ATTENTION
. If you switch off the device and there is still pressure in the system, it
is possible that the piston holder will be pushed back. As a result, the piston may be
pressed out of the syringe and the dosed medium may escape.
Therefore, relieve the pressure in the system before switching it off.
In the event of an unexpected voltage drop (e.g. power failure), take appropriate
measures to prevent push-back (e.g. check valve).
IMPORTANT
.
CETONI does not give any guarantee that any syringe will fit into the
Nemesys S universal syringe holder.
The accuracy of the dispensing depends on the correct configuration of the syringe.
Furthermore, especially in the case of plastic syringes, deviations in dosing accuracy
may occur due to deformation.
Use syringes that are approved for the expected pressure.
5