CETONI Nemesys S Hardware Manual
29
5.9
Overload shut-off
The CETONI Nemesys S has a nominal thrust force of 400 N, which enables pressure to be built up in the
syringe as well as the associated application. To protect the device and the application from damage due
to overload, it is equipped with a force measuring device. If the nominal force of 400 N is exceeded by
20%, the drive is switched off. If in this case the pressure and thus the force does not reduce on its own,
the error condition must be rectified manually. The description of this can be found in the software
manual under "What to do after a force overload stop".
In addition, you can define a second lower limit yourself. The procedure is explained in the software
manual and the Nemesys Pump Firmware Specification.
ATTENTION
. For very small syringes, pressures that are critical for the application
and the syringe can already be generated at forces that are so low that they cannot be
reliably detected by the force measuring device.
5.10
Changing the syringe holder
If a Contiflow valve is mounted, remove the connector from the accessory port as described in section
5.7. Then loosen and remove the two fastening screws with a 3 mm Allen key. Now you can take off the
syringe holder easily.