4.11 Preheating time
Preheating time
13 seconds
Change stabilisation time in
environmental conditions
5 minutes
NOTE
: The preheating time for the STRUCTURE application is 15 seconds whenever more than one
minute has elapsed since the equipment was switched on.
4.12 Influence of temperature
Operating range
-10 to +50 °C
Storage without batteries
-20 to +60 °C
4.13 Influence of humidity
Operating range
25 to 90 %
4.14 Electromagnetic compatibility
Set of accessories tested in the
verification of the
electromagnetic compatibility
requirements
All those described in the section 4.21
Operating mode and position that
produces the maximum emission
of radiofrequency
Normal* operating mode of the vibrometer
according to the accelerometer used
(AC031, AC032 or AC033),
VC431
in the
upright position, with the main axis of the
equipment perpendicular to the direction
of propagation of the field with vertical
polarisation. With all the cables connected
and powered through the PC.