9.4.6 Material tests and automatic repetitions
Material tests can be activated wit checkboxes in the settings. To activate or deactivate
Material Test Mode, navigate to the Settings Menu by clicking on the
symbol at the
top right of the screen. Press on “Sterilization” and then “Material testing mode”.
If Material Test Mode has been activated, the autoclave asks how often the program
is to be executed automatically in succession before the program start.
You can enter a number between 1 and 1000. The program will execute the desired
number of cycles fully automatically in succession. The status of the program cycle
can be called up at any time via a browser at
www.certocloud.com
or offline on the
CertoClav server at
http://certocloud
. If you have activated email or SMS notifications,
you can call up the current status of the material test by SMS or email. The email or
SMS notifications can be configured in the settings. You can find the notifications in
the Settings under the Notifications menu item.
9.4.7 Status screen
The Status screen becomes visible as soon as a program is running.
Information on the current cycle is listed on the left of the screen. The current steam
temperature and data on the pressure are on the right of the screen. If the liquid