Instruction Manual CertoClav Vacuum Pro 45 l — Model STE-45-T
13
6 Installation
6.1 Setting up the autoclave
If you wish to place the autoclave on a laboratory table, check whether the table is
suitable for the weight of the autoclave.
For problem-free operation, the autoclave must be aligned by adjusting the device feet
using spirit level in direction inclined backwards so that condensation can drain off in
the drainage port at the rear end of the autoclave. The autoclave should be installed
horizontally in a vertical direction.
Please leave at least 10 cm free space around the
autoclave. Leave at least 20 cm free space at the ventilation
slits on the right of the autoclave. Make sure that the room
is well ventilated
Keep the specified distance to the surrounding areas.
Failure to comply can lead to heat accumulation. This could
impair the function of the autoclave and shorten the service
life of the pressure pump and result in prolonged program
times.
Connect the power cable to the device socket on the rear of the autoclave.
Note:
Check whether the data on the type plate of the autoclave correspond to the
data for the power source. Check whether the safety switch of the autoclave is in the
upper position. Switch on the autoclave at the rocker switch.