10.
Care and Maintenance
To ensure safe and hygienic operation, please observe the following:
⚫
To protect the TESLA 21+ LASER SHOWER, in particular to avoid damage to the
protective glass/the comb tips, store it e.g. in the case that can be purchased
optionally.
⚫
The laser beam outlet must be cleaned with a soft cloth/cotton using 70 %
ethanol (medical alcohol) every time it has come in contact with skin or when
changing patients.
⚫
Do not expose the unit to direct sunlight.
⚫
The housing may be cleaned with a mild detergent. Before you start cleaning,
disconnect from the Battery.
⚫
Unauthorized repairs or modifications of the unit may put users and clients in
jeopardy. Therefore, repairs may only be performed by the manufacturer or by
persons who have been authorized by him. Illicit opening will void the warranty.
When repair is required, please contact the Service Department.
⚫
The laser shower should only be used at moderate temperature.
11.
Safety information
Please strictly observe the instructions in the operating manual and heed the
warnings.
⚫
To protect the TESLA 21+ LASER SHOWER, in particular to avoid damage to the
protective glass/the comb tips, store products in the case.
⚫
The laser shower must be protected against unauthorized users with the key
pin.
⚫
The TESLA 21+ LASER SHOWER must only be operated by specialized personnel,
that have been trained in handling the unit. Treatment must be supervised at all
times.
⚫
Due to its wavelength, laser radiation is not or only very weakly visible.
Do
not look into the laser beam outlet and do not point the laser beam into the
eyes of other persons, even if the eyes are closed.
⚫
The safety distance when looking straight into the beam at maximum output
(21 x 100 mW
–
CW)
is 80 inches (200 cm). If used incorrectly, lasers can
damage the eyes, especially if the distance between the eye and the laser beam
outlet is shorter than the safety distance.