page 13 of 30

 

 

Press the SET key to confirm the value. Now the required therapy time is flashing. 
It can be confirmed with the SET key.

 

 

 

Possible settings for the therapy time are:

 

10 seconds to 1 minute in increments of 10 seconds 
1 minute to 10 minutes in increments of 1 minute

 

 

Possible settings for the dose are:

 

1 J to max. 60 J

 

 

 

Please note:

 

The  maximum  energy  dose  rate  that  can  be  selected  is  determined  by  the 
maximum therapy time and the maximum laser power level of TESLA 21+ LASER 
SHOWER.

 

Example

:  Maximum therapy time for the TESLA 21+ LASER SHOWER (100 mW) is 

10 minutes. The maximum energy dose rate that can be set is therefore 100 mW 
x 10 minutes x 60 s = 60 J/LD. The minimum dose that can be selected is 1 J.

 

 

 
 

Note

:

 

Since energy and therapy time are correlated, changing either one of 
the two parameters will impact the other parameter. 

     The other value will be adjusted automatically.

 

 

Please note: 

Changing the therapy time with the 

“+” or “

-

-key also changes 

the

 

dose needed for set therapy time depending on the selected output.

 

Attention:

 When all laser parameters have been set, this is shown in the 

display. The

 

individual values are no longer flashing, and you may start with 

the laser treatment (● section 7.0 *Using the laser*, pages 14ff).

 

Summary of Contents for TESLA 21+

Page 1: ...Operating Manual For Veterinary use only Last revised June 12 2020...

Page 2: ...pson of the electron During that period in the mid 1890 s Tesla conversed often with Thompson particularly in the electrical journals At about the year 1918 Tesla apparently had a laser like apparatus...

Page 3: ...by Coleman Czito s grandson s wife it is very possible that Tesla used the same or similar kind of apparatus to send laser pulses to the moon Working of Ruby Laser 1 A powerful lamp is wrapped around...

Page 4: ...e action Copyrights The Operating Manual are subject to copyright law All rights are reserved Copying duplicating translating and converting into an electronic format in whole or in part are prohibite...

Page 5: ...ns 19 12 1 Definition of symbols of the Centurion TESLA 21 LASER SHOWER 20 13 Error messages and safety information 21 14 Technical data The Centurion TESLA 21 LASER SHOWER 22 15 Appendix Definition o...

Page 6: ...l as dose 6 LEDs 6a Two yellow LED On when laser output active Normal Operation 6b Two red LED On as error indication of LED diode operation Please contact customer service 7 red connector Connector f...

Page 7: ...ER SHOWER casing top side 6b 6a 10 11 9 8 7 5 3 fig 2 2 shower adapter top fig comb adapter lower fig 1 See Appendix 1 2 3 PM11 8800 Li ion Battery Pack For Centurion TESLA 21 LASER SHOWER fig 3 AC ad...

Page 8: ...nd can be used for helping wound healing and relief of pain and inflammation 2 Contraindications Theprofessionalliteratureliststhefollowingcontraindications Theyalsoapplyto the TESLA 21 LASER SHOWER A...

Page 9: ...J LD are displayed Factory setting whendeliveringis 100mW Multi 5min 15J LD Avaluethatappearsinvertedcanbechangedwiththe and keys Onthisscreenthelaserpowerlevelcanbe altered Pressing the SET key saves...

Page 10: ...ht alignment If you reconnect the TESLA 21 LASER SHOWER with the AC adapter make sure you observe the direction of the arrows on the plugs 3 2 Connect AC Adapter to the power supply Connect the AC ada...

Page 11: ...hown the following initial settings The maximum output 100 mW the operation mode Multi and an energy dose of 60 J LD LD laser diode are preset when operated for the first time Following this the laser...

Page 12: ...alue in increments of 10 Press the SET key again and the selected value is confirmed and stored The display changes showing the following information The therapy program can be changed with the and pa...

Page 13: ...d and isflashing 3 Select the FI Bands using the or key The display now shows FI xxxxx 4 Press the arrow key for 2 3 seconds The display now shows in the first line the frequency FI autorun and in the...

Page 14: ...to the treatment screen on the display at first as a flashing value The frequency is stored with the SET key and displayed on the screen A set value remains in the memory until it is overwritten by an...

Page 15: ...py time for the TESLA 21 LASER SHOWER 100mW is 10 minutes The maximum energy dose rate that can be set is therefore 100 mW x 10 minutes x 60 s 60 J LD The minimum dose that can be selected is 1 J Note...

Page 16: ...e emitted energy dose rateinJoule The endof the first line displays the mode selected The selected therapy time now counts backwards and the energy dose rate changes in line with the selected laser po...

Page 17: ...emperature of 55 the mode is switched automatically to multi frequency mode The temperature symbol on the display will inform you page 21 During laser emission mini fans will see to an improved remova...

Page 18: ...ion fig 9 2 In this position you can remove the adapter fig 9 3 When attaching the adapter make sure that the detents are inserted into the proper recesses Press the adapter gently against the housing...

Page 19: ...ed please contact the Service Department The laser shower should only be used at moderate temperature 11 Safety information Please strictly observe the instructions in the operating manual and heed th...

Page 20: ...tment Magnetic and electric fields as well as ionizing radiation may influence the functioning of the unit For this reason you should not operate near devices which generate large electromagnetic fiel...

Page 21: ...igits 5 6 starting from the right year of manufacture 20 2020 Manufacturer s nameplate This warning sign marks the laser beam outlet of the unit Do not look into the laser beam outlet without laser pr...

Page 22: ...and must not be disposed of in the domestic waste Distributor 12 2 Definition of symbols on AC adapter Manufacturer date only AC adapter in form YY MM Device of laser protection class II only AC Adapt...

Page 23: ...on Laser output toohigh Whenthismessageisdisplayed theoutputemitted varies from the output selected Laser output still can continue This message may also occur due to reflection of laser light on wet...

Page 24: ...approx 310 x 105 x 47 mm Total weight 430 g Classification Protection class II according to EN 60601 1 Application unit BF according to EN 60601 1 Without moisture protection IPX0 Laser class 3B acco...

Page 25: ...cies Display Frequeny Multi Frequency 200 Hz to 3 5 kHz random signal derived from the quartz oscillator impulse duration max 0 25 msec Alpha Frequenz 10 Hz Fi Band Regeneration Pain Trauma Detox Infe...

Page 26: ...s that the device meets the EMC requirements only if the accessories are used which are listed in the EC Declaration of Conformity Using other accessories may result in an increased emission of electr...

Page 27: ...nly for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions Class B The TESLA 21 LASER SHOWER laser s...

Page 28: ...3 2 0 5 kV 1 kV 0 5 kV 1 kV Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage vari ations on power supply input lines IEC...

Page 29: ...MHz 5 7GHz 9 28V m NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and ref...

Page 30: ...ment parts Freight excluded Any repairs or service work done by an unauthorized person or company will void the Warranty immediately An Extended Warranty Package is available to cover any problems tha...

Page 31: ...VE EDISON N J 08837 PHONE 1 732 417 0500 PHONE 1 800 387 8326 Email info centurionsystems com Delivered by ELEC WESTERN MEDICAL DEVICES LTD 1015 Matheson Blvd East Suite 8 Mississauga On L4W 3A4 PHONE...

Reviews: