page 13 of 30
Press the SET key to confirm the value. Now the required therapy time is flashing.
It can be confirmed with the SET key.
Possible settings for the therapy time are:
10 seconds to 1 minute in increments of 10 seconds
1 minute to 10 minutes in increments of 1 minute
Possible settings for the dose are:
1 J to max. 60 J
Please note:
The maximum energy dose rate that can be selected is determined by the
maximum therapy time and the maximum laser power level of TESLA 21+ LASER
SHOWER.
Example
: Maximum therapy time for the TESLA 21+ LASER SHOWER (100 mW) is
10 minutes. The maximum energy dose rate that can be set is therefore 100 mW
x 10 minutes x 60 s = 60 J/LD. The minimum dose that can be selected is 1 J.
Note
:
Since energy and therapy time are correlated, changing either one of
the two parameters will impact the other parameter.
The other value will be adjusted automatically.
Please note:
Changing the therapy time with the
“+” or “
-
“
-key also changes
the
dose needed for set therapy time depending on the selected output.
Attention:
When all laser parameters have been set, this is shown in the
display. The
individual values are no longer flashing, and you may start with
the laser treatment (● section 7.0 *Using the laser*, pages 14ff).