STEP 3:
Thread the stylus assembly into the stylus mount boss on the DP-7. Insert a
wrench into the hole located closest to stylus mount boss. Place the 3/16" wrench, on the
flats of the stylus mount boss. Refer to the pictures below.
Use thumb and finger to apply torque to the hex wrench. Do not apply any twisting force with
the 3/16" open end wrench; use the wrench to counter act the twisting force being applied by
the hex wrench in the hole. Excessive twisting force applied to the probe through the stylus
mount boss could damage the internal sensor assembly! If the 035” hex wrench begins to
bend too much force is being applied.
RUN-OUT ADJUSTMENT:
The run-out adjustment procedure, is necessary to ensure that the center of the probe stylus
tip is aligned with the center of the spindle rotation axis. This eliminates any unexpected off-
set when digitizing or doing automated part set-up.
STEP 1:
Position the dial indicator, as shown in the photo below, with the finger of the dial
indicator on the front and center of the stylus ball.