STEP 5: Repeat steps 4 until no run-out of the tip is visible on the dial indicator needle. If a set screws become too tight
before you have finished the adjustment, do not force them, instead loosen the setscrew that corresponds with the highest
reading on the dial indicator, and tighten the set screw at the lowest reading. Only adjust a set screw when it is directly
above the indicator finger. This way you can see the full effect of the adjustment you are making. The final small adjust
-
ment is best made by loosening a set screw at the high point of run-out.
Turning the set screw Clockwise moves the
screw in,
Causing the probe body to move out.
Turning the set screw Counter Clockwise
moves the screw out,
causing the probe body to move in.