Perform Method
WARNING
CEM Recommends that a post-method cool down time be used for each digestion performed in the
MARS 6 instrument to prevent the possibility of operator burns or acid spills
Proper precautions must be taken to avoid contact with reagents or reagent vapors. Protective gear
should be worn as outlined in the user’s safety program for hazardous materials and the reagent
manufacturer’s material safety data sheet. Refer to these guidelines for proper handling and disposal of
the reagent.
ReactiGuard alerts the operator to the occurrence of an event within a vessel; therefore, CEM
recommends that ReactiGuard be “on” at all times.
Use extreme caution while attempting to identify the underlying cause(s) of a detected vessel event.
Always wear protective gear such as gloves, a lab coat, eye protection, etc.
CAUTION
To lengthen the lifetime of the display, CEM recommends that a screen saver time be used at all times.
1. Prepare the vessels selected for the digestion in accordance with procedures outlined in the vessel
manual.
Note:
Minimum volume for the MARS 6 cavity is 10 mL of acid or 50 mL of water.
2. From the MARS 6 home screen, select either “One Touch Methods” or “Classic Methods.”