5
Installation
The MARS 6 instrument should be installed on a laboratory work bench with access to a fume hood or other means of
fume disposal.
Tools Required
Lab Coat
Gloves
Eye Protection
Installation Site
To install the MARS 6, choose a location that:
provides at least 8 in. (20 cm) space on each side and 6 in. (15 cm) space in the rear for proper ventilation.
is free from vibration of large equipment and/or walk-through traffic.
is away from the primary laboratory exits and walk-through traffic.
provides a temperature range of 41°F (5°C) to 104°F (40°C).
provides adequate bench space for sample handling and printer placement (if applicable).
permits the instrument to be connected to a dedicated, grounded outlet. The MARS 6 instrument should be
operated on stabilized, constant voltage AC power supply. (Refer to Instrument Specifications in this manual.)
Note:
Measure line voltage to ensure that it meets system specifications.
CAUTION
Line voltage fluctuations greater than 10% will affect instrument performance.
provides access to a fume hood or other means of fume disposal.
CAUTION
Never install an MARS 6 system inside a laboratory fume hood. Acid and chemical fumes may attack electrical
components, resulting in possible damage and malfunctioning of the cavity door safety interlocks.
NOTE
The MARS 6 system is supplied with eight (8) ft. of flexible exhaust hose (3-1/8” OD x 2-13/16” ID). The
inboard exhaust mechanism will push fumes through the hose at 125 CFM. To properly vent the MARS 6, the
fume hood or exhaust line must draw at least 125 CFM at the point of connection. The system must be located
within 8 feet of a hood or exhaust line for proper ventilation.